Calibration Command Reference
3-12
:SWITch?
(CALibration:PROTected:LLEVel:SWITch?)
Purpose
To request the state of the low-level calibration lock.
Format
:cal:prot:llev:swit?
Response
0
Low-level calibration locked
1
Low-level calibration unlocked
Description
:SWITch? query requests the status of the low-level calibration lock from the instrument. This
:SWITch? query should not be confused with the :SWITch? query that requests the status of the
comprehensive calibration lock (see paragraph 3.3.2.)
Programming Note
To unlock low-level calibration, hold in the CAL switch while turning on instrument power.
:CAL:PROT:LLEV:SWIT?
Request low-level CAL switch status.
:STEP (CALibration:PROTected:LLEVel:STEP)
Purpose
To program individual low-level calibration steps.
Format
:cal:prot:llev:step <n>
Parameters
1
20V AC @ 1kHz
2
20V AC @ 30kHz
3
200V AC @ 1kHz
4
200V AC @ 30kHz
5
1.5V AC @ 1kHz
6
200mV AC @ 1kHz
7
5mV AC @ 100kHz
8
0.5mV AC @ 1kHz
9
+100V DC
10
-20V DC
11
Rear inputs short-circuit.
12
20mA AC @ 1kHz
13
2V AC @ 1HZ
Description
The :STEP command programs the 13 individual low-level calibration steps; <n> represents the
calibration step number. The appropriate signal must be connected to the instrument when pro-
gramming each step, as summarized in the parameters listed above (see Section 2 for details).
Example
:CAL:PROT:LLEV:STEP 7
Program low-level step 7.
Содержание 2002
Страница 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 9: ...iv...
Страница 11: ...vi...
Страница 55: ...Calibration 2 26...
Страница 70: ...Specifications A 1 A Specifications...
Страница 83: ...A 14 Specifications...
Страница 101: ...Calibration Programs B 18...
Страница 113: ......