Calibration Command Reference
3-11
:A2
(CALibration:PROTected:DC:A2)
Purpose
To program the 2A comprehensive calibration step.
Format
:cal:prot:dc:a2 <Cal_current>
Parameter
<Cal_current> = 0.95 to 2.05 [A]
Description
A2 programs the 2A comprehensive calibration step. The allowable range of the calibration cur-
rent parameter is from 0.95A to 2.05A. Use the 1A value whenever possible for best results.
Example
:CAL:PROT:DC:A2 1
Program 2A step.
3.3.9 :LLEVel
(:CALibration:PROTected:LLEVel)
Low-level calibration commands are summarized in Table 3-3.
Table 3-3
Low-level calibration commands
Command
Description
:CALibration
:PROTected
:LLEVel
:SWITch?
:STEP <Step #>
1
2
3
4
5
6
7
8
9
10
11
12
13
Low-level calibration subsystem.
Request low-level CAL switch state.
(0 = locked; 1 = unlocked)
20V AC at 1kHz step.
20V AC at 30kHz step.
200V AC at 1kHz step.
200V AC at 30kHz step.
1.5V AC at 1kHz step.
0.2V AC at 1kHz step.
5mV AC at 100kHz step.
0.5mV AC at 1kHz step.
+100V DC step.
-20V DC step.
Rear inputs short-circuit step.
20mA AC at 1kHz step.
2V AC at 1Hz step.