Calibration
2-5
Procedure
Step 1: Prepare the Model 2002 for calibration
1. Turn on the power, and allow the Model 2002 to warm
up for at least four hours before performing calibration.
2. Unlock comprehensive calibration by briefly pressing in
on the recessed front panel CAL switch, and verify that
the following message is displayed:
CALIBRATION UNLOCKED
Calibration can now be performed
3. Enter the front panel calibration menu as follows:
A. From normal display, press MENU.
B. Select CALIBRATION, and press ENTER.
C. Select COMPREHENSIVE, then press ENTER.
4. At this point, the instrument will display the following
message:
DC CALIBRATION PHASE
Step 2: DC zero calibration
1. Press ENTER. The instrument will display the follow-
ing prompt.
SHORT CIRCUIT INPUTS
Table 2-3
Front panel comprehensive calibration summary
Step
Description
Equipment/
connections
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
Warm-up, unlock calibration
DC Zero calibration
+2V DC calibration
+20V DC calibration
1M
Ω
calibration
200k
Ω
calibration
20k
Ω
calibration
2k
Ω
calibration
200
Ω
calibration
20
Ω
calibration
200µA DC calibration
2mA DC calibration
20mA DC calibration
200mA DC calibration
2A DC calibration
Open-circuit calibration
AC self-calibration
Enter calibration dates
Save calibration constants
None
Low-thermal short
DCV calibrator
DCV calibrator
Ohms calibrator
Ohms calibrator
Ohms calibrator
Ohms calibrator
Ohms calibrator
Ohms calibrator
DCA calibrator
DCA calibrator
DCA calibrator
DCA calibrator
DCA calibrator
Disconnect leads
Disconnect leads
None
None
2. Connect the Model 8610 low-thermal short to the instru-
ment INPUT and SENSE terminals, as shown in Figure
2-1. Wait at least three minutes before proceeding to
allow for thermal equilibrium.
NOTE
Be sure to connect the low-thermal short
properly to the HI, LO, and SENSE termi-
nals. Keep drafts away from low-thermal
connections to avoid thermal drift, which
could affect calibration accuracy.
3. Press ENTER. The instrument will then begin DC zero
calibration. While calibration is in progress, the follow-
ing will be displayed:
Performing Short Ckt Calibration
Step 3: DC volts calibration
1. When the DC zero calibration step is completed, the fol-
lowing message will be displayed:
CONNECT 2 VDC
2. Disconnect the low-thermal short, and connect the DC
calibrator to the INPUT jacks, as shown in Figure 2-2.
NOTE
Although 4-wire connections are shown,
the sense leads are connected and discon-
nected at various points in the procedure
by turning calibrator external sense on or
off as appropriate. If your calibrator does
not have provisions for turning external
sense on and off, disconnect the sense
leads when external sensing is to be turned
off, and connect the sense leads when
external sensing is to be turned on.
Figure 2-1
Low-thermal short connections
Model 2002
S+
HI
LO
S-
Model 8610
Low-thermal
short
Содержание 2002
Страница 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 9: ...iv...
Страница 11: ...vi...
Страница 55: ...Calibration 2 26...
Страница 70: ...Specifications A 1 A Specifications...
Страница 83: ...A 14 Specifications...
Страница 101: ...Calibration Programs B 18...
Страница 113: ......