Calibration
2-14
Table 2-7
Recommended equipment for low-level calibration
Mfg.
Model
Description
Specifications*
Fluke
5700A
Calibrator
±5ppm basic uncertainty.
DC Voltage:
±2V: ±7ppm
+20V: ±5ppm
+100V: ±7ppm
Resistance:
19
Ω
: ±26ppm
190
Ω
: ±17ppm
1.9k
Ω
: ±11ppm
19k
Ω
: ±11ppm
100k
Ω
: ±13ppm
1M
Ω
: ±18ppm
DC Current:
200µA: ±100ppm
2mA: ±55ppm
20mA: ±55ppm
200mA: ±65ppm
1A: ±110ppm
AC Voltage:
0.5mV @ 1kHz: ±10000ppm
5mV @ 100kHz: ±2400ppm
200mV @ 1kHz: ±150ppm
1.5V @ 1kHz: ±80ppm
20V @ 1kHz: ±80ppm
20V @ 30kHz: ±140ppm
200V @ 1kHz: ±85ppm
200V @ 30kHz: ±240ppm
AC Current:
20mA @ 1kHz: ±160ppm
Keithley
3930A or
3940
Synthesizer
2V rms @ 1Hz
Keithley
8610
Low-thermal
shorting plug
* 90-day calibrator specifications shown include total uncertainty at specified output.