10
INITIAL SETUP INSTRUCTIONS, continued
STEP 6
–
Adjust Pressure.
A.
Place a seal on an empty vial, and hold the
vial up into the Head as far as it will go.
B.
While holding the vial in place (A), pull
the Crimp Lever (B) forward until it
stops. Gently release the Crimp Lever
and remove the vial from the Head.
NOTE:
The Crimp Lever is adjustable, so it may be
moved to a position most comfortable for
the operator. See label near Crimp Lever
for adjustment instructions.
C.
Inspect the seal on the vial.
If there is little to no crimp,
increase the pressure by turning
the Pressure Adjustment Dial (C)
clockwise by no more than 50
fine increments (half of a turn of
the dial).
If there is enough crimp to
prevent the seal from freely
moving, increase the pressure by
turning the Pressure Adjustment
Dial (C) clockwise by no more
than 10 fine increments.
If the seal is overcrimped, then decrease the pressure by turning the Pressure
Adjustment Dial (C) counterclockwise, and make smaller incremental adjustments
than previously used.
D.
Repeat steps 6.B. and 6.C. until a satisfactory crimp is achieved. Unless the seal is
overcrimped, the same vial and seal may be used for each pressure adjustment.
A
B
C