26
WARNINGS:
If the
BenchMark
is not set up properly for the vial size and seal type that is to be crimped,
there is a chance of vial breakage during the crimp cycle. As a precaution, the following
checks should be made prior to operating the
BenchMark
:
Make sure the correct Vial Holder and Head are assembled to the
BenchMark
.
Make sure all four settings are correct for the vial size and seal type to be
crimped (refer to settings recorded during the initial setup).
Wear impact resistant safety glasses and protective gloves while operating the
BenchMark
.
OPERATING INSTRUCTIONS
CRIMPING OPERATION
INSTRUCTIONS
A.
Place a seal on a filled vial and set
the vial on the Vial Holder so that it
is up against the Vial Stop (A).
B.
Pull the Crimp Lever (B) forward as
far as it will go, then gently allow
the Crimp Lever to return to its
original position.
C.
Remove the sealed vial from the
Vial Holder.
D.
Repeat steps A through C until all
vials have been sealed.
A
B