22
◆
Operation record data
・
Record file name
:
PAT05_8.csv (It will be a csv format file)
"05" is the pattern number of the operation and "_8" is the serial number of pattern 05. The serial
number is incremented each time the operation of pattern 05 is started and recorded with that file
name.
・
Record file date and time
:
This unit does not have a clock, so the dates and times of files are as follows.
2014/04/01
0:00
・
Record contents
1. At the beginning of the file, record the program content of the pattern that was operated.
2. After that, every 5 seconds, record the status of "Running segment number, current temperature,
gas output, event, AT, SS". When the operation is stopped, the contents up to the stop are
recorded.
※
For gas output, event, AT and SS, "1" is recorded during operation and "0" is recorded during
stop.
Initial Setting Mode
・
In standby mode, press the "F3" (initial setting) key to switch to this mode.
The following settings can be made in this mode.
※
UP/DOWN keys are used to change the value, and up/down keys are used to move the items.
◆
"Buzzer volume" setting
・
You can set 4 levels: mute, small, medium, and large.
1.
Key
:
Key operation sound Initial value
:
Small
2.
Error
:
Error
sound Initial
value
:
Large
3.
Others
:
Other than the above (end sound, power-on sound, etc.) Initial
value
:
Small
◆
"Temp. Calibration" setting
・
The gain of furnace temperature can be corrected for 1000
℃
.
For example, if you want to increase 20
℃
at 1000
℃
, set +20
℃
.
In that case, at 500
℃
, "20
℃×
500
℃/
1000
℃
" will be a correction of approxi10
℃
.
Setting
range
:
-50 to +50
℃
Initial
value
:
+ 0
℃