Installation of this machine 

Installation location

environment 

 

In order to operate the muffle furnace normally, please prepare the following power supply equipment 

    according to the product. 
      300Plus 

 

AC100V 50/60Hz, current capacity 20A 

      500Plus 

 

Single-phase AC200V 50/60Hz, current capacity 20A 

      1500Plus 

Single-phase AC200V 50/60Hz, current capacity 30A 

      3000Plus 

Single-phase AC200V 50/60Hz, current capacity 40A 

      5000Plus 

Single-phase AC200V 50/60Hz, current capacity 50A 

 
 

Install in a place with little dust and near the power supply facility. 

 

Install this machine on a horizontal and sturdy table, and fix it to a sturdy structure with metal wires, 

  etc., 

to 

prevent it from falling. 

 

Install the unit at a distance of 100 cm on the top and at least 30 cm away from walls and equipment. 

    Also, do not place any objects on the top surface of this machine. 
 

Wiring work 

 

For 300Plus

 

    The power cord comes out of the main unit. Connect it to the grounded outlet of the power circuit 
    prepared for this unit alone. Do not use a plug adapter, etc. There is a risk of fire or electric shock. 
 

For other than 300Plus

 

    Connect the supplied power cord to the power terminal of this machine, and then connect it to the 
    power equipment prepared exclusively for this machine.

 

The green wire of the power cord becomes 

  the 

ground 

wire. 

 
     

Be sure to connect the ground wire to a properly constructed ground terminal. 

 

Installation of hearth plate 

 

Please lay the attached hearth plate on the bottom of the furnace. 

   

For 5000Plus, lay two attached hearth plates on the bottom of the furnace. 

 

 

If you place the fired product directly on the hearth plate, the temperature may be lower 

    than the furnace temperature, so adjust it with a spacer such as refractory so that it is 
    placed in the center of the furnace as much as possible. 

The hearth plate is necessary to protect the heat insulating material inside the furnace. 

    Never use without the hearth plate. 

Since the hearth plate and the fired product may react with each other and damage 

    each other, put the fired product in a tray or crucible that does not react with the fired 
    product (alumina, mullite, etc.) before placing it on the hearth plate. 

 

Installation of exhaust guide and exhaust port plug 

 

Depending on the application, insert the attached exhaust guide or exhaust port plug into the hole near 

    the center of the furnace top surface. Insert the thin outer shape through the hole. 
 
  Exhaust 

guide 

When installing a deodorizer or smoke exhaust device, install an exhaust guide. 

  Exhaust 

plug 

 

If you are not using a deodorizer, insert the exhaust plug into the hole to close it. 

 
 

 

CAUTION 

Содержание Plus Series

Страница 1: ...tric Furnace Plus series Before using the unit please read the manual thoroughly and understand the capabilities and proper usage for this machine Please keep this manual in an easily accessible locat...

Страница 2: ......

Страница 3: ...nstallation location environment 11 Wiring work 11 Installation of hearth plate 11 Installation of exhaust guide and exhaust port plug 11 Connection of rear terminals 12 T C OUTPUT OPTION 12 CLEANER 1...

Страница 4: ...Check and change program contents 21 PID auto tuning operation 21 About SD card operation record 21 Operation record data 22 Initial Setting Mode 22 Buzzer volume setting 22 Temp Calibration setting...

Страница 5: ...on material resources Please read them carefully prior to your actual machine operation WARNING CAUTION Failure to follow or ignorance of the directions may cause severe injury or death Failure to fo...

Страница 6: ...only be used in Japan When installing this unit keep it 100 cm above the ceiling and at least 30 cm away from walls and equipment in other cases Also do not place any objects on the top surface of th...

Страница 7: ...the power switch or breaking the breaker The case of this machine will not be cooled and may cause burns Fixed the wiring and piping connected to this machine so that they will not be caught It may c...

Страница 8: ...during use but this is not a defect There is no problem with performance so use it as is Standard accessories Exhaust guide 1pc Exhaust port plug 1pc Hearth plate 1pc One set of 2 for 5000Plus AC cord...

Страница 9: ...the power Liquid crystal panel Display part of the color liquid crystal Operation panel Program operation Start processing SD card slot side Use SD card to edit programs update this unit etc Exhaust...

Страница 10: ...inel thermocouples Cleaner smoke exhaust device output Power outlet for optional cleaner and smoke exhaust device The 3000 and 5000Plus are connected using the included extension cable Do not connect...

Страница 11: ...nsulation to improve the thermal efficiency and temperature distribution For this reason if the processed product contacts the hot plate it may be damaged so be careful not to contact the left and rig...

Страница 12: ...AL2O3 protective film formed on the surface of the heating element or that reacts with the protective film to lower the melting point is harmful to the heating element S sulfur Use in an atmosphere c...

Страница 13: ...er equipment prepared exclusively for this machine The green wire of the power cord becomes the ground wire Be sure to connect the ground wire to a properly constructed ground terminal Installation of...

Страница 14: ...sting device cannot be used at the same time 300Plus Deodorizing device ES71 Smoke exhausting device VF71 500Plus 1500Plus 3000Plus 5000Plus Use the extension cable included with 3000 and 5000Plus For...

Страница 15: ...3 Course selection 2 4 Course selection 2 5 DC5 24V GND GAS UNIT OUTPUT A connector for connecting the optional gas introduction unit By connecting the gas introduction unit to this connector the gas...

Страница 16: ...rogramming the temperature control each event and each condition value The program can program up to 100 segments per pattern There are 50 patterns and it is easy to identify by entering the pattern n...

Страница 17: ...m being interrupted The fan will stop and the enclosure will become hot which is dangerous START STOP Lamp Lights during operation Also it flashes during each operation of the auto tuning and soak sto...

Страница 18: ...rn name Character English uppercase number Number of characters 16 single byte characters Input the pattern name Enterable symbols Blank Temp 0 to 1150 other than 5000Plus 0 1100 5000Plus Segment set...

Страница 19: ...egment 0 is always held In the example the operation is maintained for 0 10 at 20 Gas 1 ON is set to 0 00 so it turns on immediately after segment 0 operation after 0 00 2 Segment 1 Because the temper...

Страница 20: ...input 1 When the cursor is Seg0 press the Up key to move the cursor to the pattern name 2 Use the left right keys to move to the place you want to input and press UP or DOWN Press the key to select t...

Страница 21: ...matically selected If you want to switch to manual change the PID setting to 0 9 If set to Manual the set PID set constants are used For the setting method set the value of 0 9 or A in the PID item of...

Страница 22: ...tion of main display Seg Displays the segment number currently in progress TOTAL Time left Displays the remaining time of operation You can switch between TOTAL time and remaining time in Seg units Co...

Страница 23: ...e pattern selection screen is displayed Select the pattern you want to edit and press the enter key Changes made to patterns other than those during operation are stored in memory PID auto tuning oper...

Страница 24: ...opped the contents up to the stop are recorded For gas output event AT and SS 1 is recorded during operation and 0 is recorded during stop Initial Setting Mode In standby mode press the F3 initial set...

Страница 25: ...value SD card error detection is not performed at the start of operation Yes If there is an error at the start of operation a confirmation screen will be displayed Either this setting is set recordin...

Страница 26: ...ion of main display PID No 0 to 9 correspond to the setting number at the time of programming 10 to 19 are constants for auto setting It can be changed with the UP and DOWN keys PID temperature The te...

Страница 27: ...is mode In this mode you can read and write the program contents and write the maintenance information contents to the SD card When you want to edit the program contents on the PC back up the program...

Страница 28: ...quipment Warning content Warning No 2 Controller error Displayed when the controller operates abnormally due to the influence of noise Almost always recovers when the power is turned on again but if i...

Страница 29: ...60Hz 8 0kVA Maximum operating temperature 1150 5000 Plus only 1100 Regular use temperature 1000 Operating temperature range 100 Maximum operating temperature Note Temperature distribution and control...

Страница 30: ...e sensor abnormality Memory error over temperature error Main options Gas introduction unit 1 system Model KDG 300 Applicable models 300Plus 500Plus Flowmeter Nitrogen 2L min Oxygen argon possible 1 i...

Страница 31: ...nt duct 2 5m when extended Duct clamp Model KDF VF72 Applicable models 500Plus 1500Plus 3000Plus 5000Plus Power supply AC200V 34VA 50Hz 37VA 60Hz Accessories 75 heat resistant duct 2 5m when extended...

Отзывы: