Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
8 Troubleshooting
Malfunction
Cause
Remedy
No light on the straight or
contra-angle handpiece
The light is not turned on.
▶ Turn on the light.
See also:
2
5.2.3 Setting the LUX brightness, Page 32
The straight and contra-
angle handpiece is improp-
erly attached.
▶ Attach the straight and contra-angle handpiece un-
til the catch audibly locks.
Defective LED.
▶ Replace the LED.
See also:
2
Instructions for use INTRA LUX S600 LED
Not a suitable straight and
contra-angle handpiece.
▶ Use a suitable light, straight and contra-angle
handpiece.
Error message from software
Malfunction
Cause
Remedy
Event E3:
Release the foot pedal
Wireless foot control was
actuated while the unit
started up.
▶ Release the wireless foot control.
Event E4:
Data initialisation
Setting data newly initial-
ised.
▶ Confirm message and check or correct the pro-
gram settings.
If the problem persists, notify service engineer
Event E6:
Internal communication er-
ror
Internal system error.
▶ Turn the unit off and on. If the problem persists,
notify service engineer.
Event E9:
Setting the time and date
Time was not reset after
re-start.
▶ Setting the date and time.
▶ If the problem persists, have a service engineer re-
place the battery of the real-time clock.
Event E14:
Is shown at start-up if the
battery level is low
Battery level low.
▶ Batteries (alkaline type AA/ LR6) must be supplied
and replaced soon. Confirm the message to be
able to continue working.
Event E14:
Is shown approximately
every 10 min if battery
level is low
Battery level critical.
▶ Batteries (alkaline type AA/ LR6) must be replaced
instantaneously to be able to continue working.
Event E29:
SD card write protection
SD memory card is write-
protected or defective.
▶ Switch the write-protect switch on the left side of
the card upwards to "unlock", "write" or the like.
▶ If the problem persists, use a new SD memory
card.
▶ If the problem persists, notify service engineer.
All events > 30:
Hardware error
Internal system error.
▶ Turn the unit off and on. If the problem persists,
notify service engineer.
65 / 76