Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.8 Activating the one-touch calibration
50 / 76
5.8 Activating the one-touch calibration
The one-touch calibration automatically compensates for torque deviations of the mo-
tor that may be caused, e.g., by aging processes. When the handpiece is attached,
the unit detects if the handpiece runs sluggish or is defective. The one-touch calibra-
tion thus provides for a more accurate torque on the contra-angle handpiece.
Note
The handpiece must be attached for calibration.
One-touch calibration should be carried out only with KaVo surgical handpieces
with a transmission ratio of 16:1, 20:1 or 27:1.
The one-touch calibration cannot be carried out with third-party handpieces or
handpieces with different transmission ratios.
The calibration must be repeated whenever the handpiece is changed.
▶ Touch the calibration symbol in order to start the One-Touch calibration.
The display shows "Press wireless foot control".
CAUTION
The motor starts at full speed.
Risk of injury.
▶ Hold the motor firmly or put it in a safe holder during the calibration.
▶ Press the wireless foot control and hold it down until the display shows that "Calib-
ration was successful".
▶ If you release the wireless foot control before the display shows that the calibra-
tion was successful, press the wireless foot control again until the display shows
that the calibration was successful.
▶ Press the back key to terminate the calibration and to return to the selection of
device settings.
If an unsuitable or defective handpiece was used in the calibration, the calibration is
discontinued and the error message, "Measurement failed undefined error", is shown.
▶ Press the back key to terminate the failed calibration.
See also:
2
8 Troubleshooting, Page 64