Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
8 Troubleshooting
64 / 76
8 Troubleshooting
Note
If malfunctions cannot be located or eliminated using this troubleshooting guide, a
technician trained by KaVo must be commissioned to eliminate the problem.
Note
The permitted work is described in the Technician's Instructions available to the
trained service staff.
In case of malfunction, the display names the malfunction directly or shows an error
number.
Malfunction
Cause
Remedy
Non-functional device.
The unit is switched off.
▶ Switch-on the mains switch on the rear of the unit.
Neither end of the power
cable is plugged in.
▶ Plug in the power input cable.
Unknown.
▶ Turn the unit off and on.
Blown fuse.
▶ Contact customer service.
No coolant in the hand-
piece.
No coolant flow pre-selec-
ted.
Pump is off.
▶ Pre-select coolant flow.
See also:
2
5.6.3 Setting the coolant flow, Page 46
Hose fixation is not
plugged in.
▶ Plug in the hose fixation.
See also:
2
4.5 Connecting the coolant container and hose set,
Page 24
Bottle is empty.
▶ Connect new bottle.
Hose clamp is closed.
▶ Open the hose clamp.
Pump locking mechanism
is not closed.
▶ Check and close, if needed, the closing mechan-
ism.
Hose is kinked.
▶ Check hose and remove the kink, if any.
Glass bottle containing the
coolant is not ventilated.
▶ Open the cap on the ventilation valve of the punc-
ture needle.
Insufficient coolant flow in
the instrument.
Spray nozzles are crusty
or soiled.
▶ Clean the spray nozzles with the nozzle needle or
re-process the part.
See also:
2
SURGmatic Instructions for use
Glass bottle containing the
coolant is not ventilated.
▶ Open the cap on the ventilation valve of the punc-
ture needle.
The motor makes a grind-
ing noise or does not run
smoothly.
The motor is not correctly
plugged on or screwed on.
▶ Firmly insert the motor hose into the housing.
▶ Firmly screw on the motor hose to the motor.
▶ Check if all the connections and couplings are
firmly seated.