Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
7 Reprocessing steps in accordance with DIN EN ISO 17664 | 7.2 Disinfection
60 / 76
7.1.2 Machine cleaning
CAUTION
Damage by liquids.
Faults on electrical components.
▶ Do not machine-clean the device.
The following parts of the unit are released for machine-based cleaning:
▪ Handpiece tray
▪ Motor and motor cable
Note
Please comply with the corresponding Instructions for use when you re-process the
motor and motor cable.
The re-processing steps for the handpiece tray are as follows:
KaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that are
operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD program, the cleaning agent
neodisher® MediClean and the neutralisation agent neodisher® Z.
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector.
7.2 Disinfection
Note
After each treatment of a patient, the surfaces near the patient that may have been
contaminated by contact or aerosol need to be disinfected. All disinfection meas-
ures need to be carried out by wipe disinfection.
Note
Please comply with the corresponding Instructions for use when you re-process the
motor and motor cable.
7.2.1 Manual disinfection
▶ Use a soft disposable cloth and an approved disinfectant for disinfection by wiping
down all visible surfaces of the unit, bottle holder, wireless foot control surfaces,
and connecting cables. Make sure that all surfaces are wetted.
▶ Let the disinfectant act for the prescribed time.
▶ Dry the surfaces.
Permissible disinfectants (application range in accordance with the available manufac-
turer's instructions and national guidelines. Please note material safety data sheets.)
KaVo recommends the following products based on the compatibility of the materials.
The microbiological efficacy must be confirmed by the disinfectant manufacturer.
▪ FD 322 Dürr
▪ Microcide AF Liquid (Schülke & Mayr)
▪ CaviCide made by Metrex