Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
4 Commissioning | 4.6 Electrical connection
▶ Stick the puncture needle ⑥ into the coolant container and hook-in the coolant
container on the bottle holder.
▶ Check the sealing and firm seating of the puncture needle ⑥. Prevent fluid from
leaking above the device.
▶ If you use a glass bottle, open the ventilation on the puncture needle ⑥.
▶ If you use a bag, keep the ventilation on the puncture needle ⑥ closed.
▶ Open the hose clamp ⑦ before startup.
Note
Using a new hose, it may take up to approx. 10 seconds for the coolant to exit on
the handpiece, depending on the feed rate.
4.6 Electrical connection
CAUTION
Damaged mains cable / missing protective conductor.
Electrical shock.
▶ Check the mains cable before use. The socket outlet must have a protective con-
tact and meet the respective national guidelines.
29 / 76