Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.5 Setting and executing program steps
Programme
step
Icon
Activity
Speed [rpm]
Torque [Ncm]
Transmission
ratio
Coolant flow
3
Template
drilling
200 – 2,000
500 (D)
5 – 20
10 (D)
16:1
27:1
20:1 (D)
0 – 4
2 (D)
4
Tapping
15 – 50
20 (D)
5 – 80
25 (D)
16:1
27:1
20:1 (D)
0 – 4
2 (D)
5
Placing im-
plant
15 – 50
20 (D)
5 – 80
25 (D)
16:1
27:1
20:1 (D)
0 – 4
0 (D)
6
Setting a clos-
ure cap
15 – 50
20 (D)
5 – 15
8 (D)
16:1
27:1
20:1 (D)
0 – 4
0 (D)
7
Free use
300 – 40,000
40,000 (D)
0.15 – 5.5
3 (D)
1:1
0 – 4
2 (D)
20 – 2,000
5 – 80
16:1
15 – 2,000
5 – 80
20:1
15 – 1,200
5 – 80
27:1
8
Rinsing func-
tion
–
–
–
–
9
Treatment
completed
(can be set
from pro-
gramme step
4)
–
–
–
–
(D) = factory setting (default setup)
Note
The listed indications are only examples. In order to prevent risks, it is essential to
comply with the manufacturer recommendations concerning implants, handpieces,
and tools.
41 / 76