Signal paths
Figure 18
Measuring sequences
During flow measuring, ULTRAFLOW® passes through a number of sequences which are repeated at fixed
intervals. Deviations only occur when the meter is in test mode and when connecting the supply during
initialization/start-up. The routines of normal mode are listed in the table below.
Time [s] Operation
0
Phase difference and absolute time measurement with and
against the flow as well as pulse emission
1
Pulse emission
2
Pulse emission
3
Phase difference and absolute time measurement with and
against the flow, reference measurement and pulse
emission
4
Pulse emission
5
Pulse emission
6
Phase difference and absolute time measurement with and
against the flow as well as pulse emission
7
Pulse emission
8
Pulse emission
9
Phase difference and absolute time measurement with and
against the flow as well as pulse emission
10
Pulse emission
11
Pulse emission
12
Phase difference and absolute time measurement with and
against the flow as well as pulse emission
Table 7
The routines are the same in test mode, but with 1 sec. intervals between measurings instead of 3 sec. as
in normal mode.
It may take up to 16 seconds to obtain correct function after a power failure.
Q
3
:
. … ³/h
Triangle
The sound path covers the measuring
pipe in a triangle, sent from the
transducers around the measuring
pipe via reflectors.
Q
3
: 1.6 m³ h
2 parallel tracks
The sound path is parallel with
the measuring pipe, sent from
the transducers via reflectors.