Page 30
Instruction manual - G95
Kivex Biotec Ltd
11. Trouble shooting
SYMPTOM
CAUSE
ACTION
Temperature alarm is on.
Temperature more than
+/- 0,5° C from set tem-
perature.
Check the set tempera-
ture. Wait for the tem-
perature to stabilise.
No heating.
Setpoint below ambient.
Change setpoint
(eg. 37° C).
Display remains off when
“HEAT” key is activated.
No mains supply.
Plug in the power
connector.
Power connector,
not connected.
Replace the display PCB.
Defective “HEAT” key.
Replace the keyboard.
Display read "E-1".
Defect temperature sen-
sor.
Replace temperature sen-
sor.
No heating.
Setpoint below ambient.
Change setpoint
(eg. 37° C).