Page 28
Instruction manual - G95
Kivex Biotec Ltd
9.6. Purging
Attach the gas supply to the rear of the in-
cubator, where the gas connector is
situated.
Note
Make sure to read all instructions regarding the gas supply.
When gassing, the lid should be closed but
not secured with the locks, in order for the
air to exit the incubator.
When the portable incubator is opened, the gas mixture is partly replaced with atmospheric air.
The atmospheric air mixture must be replaced by the correct gas mixture as quickly as possible.
The flowrate should be high for 20-30 seconds after opening and then set back to the lower value.
When frequent openings of the portable incubator are required, it may be necessary to use a
flowmeter with a flush-flow valve for rapid purging. In this case, K-SYSTEMS recommends the GH01.
After gassing, the lid should then be se-
cured with the two locks and the incubator
is ready for transport.