10. User maintenace
10.1 Cleaning
The G95 is a low maintenance unit designed to be both easy to use and clean, plus it is durable due to its
construction.
All surfaces should be cleaned with a 70� alcohol solution on a clean cloth or lint-free paper towel.
Clean all surface components by wiping gently with gauze.
•
To remove fingerprints or oil smudges, wipe with gauze slightly moistened with a mixture
•
of ether (70%) and alcohol (30%).
Note
Since solvents such as ether and alcohol are highly flammable, they must be handled
carefully. Be sure to keep these chemicals away from open flames or potential sources
of ignition – for example, electrical equipment that is being switched on or off.
Also remember to always use these chemicals only in a well ventilated room.
If smudges are difficult to clean, wipe them with a soft cloth slightly moistened with a
•
diluted neutral detergent.
The heated compartment is able to withstand some water spills. However, precautions are
•
still necessary if water is spilled on the surface. Unplug the power cord, and then wipe dry
with a dry cloth immediately.
Note
If water gets inside the unit, contact your K-SYSTEMS representative to check electrical
safety.
10.2 Calibration
For optimal performance the unit should be calibrated once a year by authorized K-SYSTEMS Service
provider.