FP51-SL
89
Command
Parameter
Response of circulator
IN_PAR_13
none
Adjusted maximum internal temperature of the cascade
controller.
IN_PAR_14
none
Adjusted minimum internal temperature of the cascade
controller.
IN_PAR_15
none
Band limit (upper)
IN_PAR_16
none
Band limit (lower)
IN_MODE_01
none Selected
setpoint:
0 = Setpoint 1
1 = Setpoint 2
2 = Setpoint 3
IN_MODE_02
none Selftuning
type:
0 = Selftuning “off”
1 = Selftuning “once”
2 = Selftuning “always”
IN_MODE_03
none
Type of external programmer input:
0 = Voltage 0 V to 10 V
1 = Current 0 mA to 20 mA
IN_MODE_04
none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
IN_MODE_05
none
Circulator in stop/start condition:
0 = stop
1 = start
IN_MODE_08
none
Adjusted control dynamics
0 = aperiodic
1 = standard
IN_HIL_00
none
Max. cooling power (%).
IN_HIL_01
none
Max. heating power (%).