FP51-SL
87
Command
Parameter
Response of circulator
OUT_MODE_03
1
Set external programmer input to current.
Current
0 mA ... 20 mA
OUT_MODE_04
0
Temperature control of internal bath.
OUT_MODE_04
1
External control with Pt100 sensor.
OUT_MODE_05
0
Stop the unit = R –OFF-.
OUT_MODE_05
1
Start the unit.
OUT_MODE_08
0
Set the control dynamics - aperiodic
OUT_MODE_08
1
Set the control dynamics - standard
OUT_SP_00
xxx.xx
Set working temperature. “Setpoint 1”
OUT_SP_01
xxx.xx
Set working temperature. “Setpoint 2”
OUT_SP_02
xxx.xx
Set working temperature. “Setpoint 3”
OUT_SP_03
xxx.xx
Set high temperature warning limit “OverTemp”
OUT_SP_04
xxx.xx
Set low temperature warning limit “SubTemp”
OUT_SP_06
xxx.xx
Set manipulated variable for heater via serial interface
-100 ... +100 [%]
OUT_SP_07
x
Set the pump pressure stage. (1 ... 4)
OUT_PAR_04
x.x
CoSpeed for external control 0 ... 5.0.
OUT_PAR_06
xxx
Xp control parameter of the internal controller.
0.1 ... 99.9
OUT_PAR_07
xxx
Tn control parameter of the internal controller.
3 ... 9999
OUT_PAR_08
xxx
Tv control parameter of the internal controller.
0 ... 999
OUT_PAR_09
xxx
Xp control parameter of the cascade controller.
0.1 ... 99.9
OUT_PAR_10
xxx
Proportional share of the cascade controller.
1 ... 99.9
OUT_PAR_11
xxx
Tn control parameter of the cascade controller.
3 ... 9999
OUT_PAR_12
xxx
Tv control parameter of the cascade controller.
0 ... 999
OUT_PAR_13
xxx
Maximum internal temperature of the cascade controller.
OUT_PAR_14
xxx
Minimum internal temperature of the cascade controller.
OUT_PAR_15
xxx
Band limit (upper)
0 ... 200 °C
OUT_PAR_16
xxx
Band limit (lower)
0 ... 200 °C
OUT_HIL_00
-xxx
Set the desired maximum cooling power (0% to 100%).
Note:
Enter the value with a preceding negative sign!
This setting is meaningful only for FP cooling machines.
OUT_HIL_01
xxx
Set the desired maximum heating power (10% to 100%).