Menu functions
64
Î
x
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
Return to 2. Set working temperature value: 120.00 °C
Î
The value >TmpValue< is only indicated.
L
In addition the measured temperature
value >CalValue X< is saved during the
next step.
Î
x
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
Return to 2. Set working temperature value: 160.00 °C
Î
The value >TmpValue< is only indicated.
L
In addition the measured temperature
value >CalValue X< is saved during the
next step.
Î
x
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
5. Return to standard operation
Î
Set >YES< after calibration.
(Standard operation)
x
The parameter flashes, switch by
pressing
and