Menu functions 

50 

9.4.5. 

ActVar - actuating variable 

 

 

 

 

 

 

 

 

The variable 

(ACTuating VARiable) 

corresponds to the extent to which 

the heater or cooling unit of the circulator is controlled. Heat or cold is 
applied to the bath according to this variable. If this happens with the 
control electronics of the circulator, called > control < in this particular 
case, the bath temperature is exactly heated and maintained constant at 
the adjusted setpoint. 

L

 

Programming of variables for the parameters > SERIAL < or > 
EPROG < is only accepted, if the unit is in Start mode. 

 

Possible parameters: 

control

 – The internal control electronics of the circulator controls the 

heater and the connected cooling unit. Self-tuning is possible.. 
(factory setting) 

serial

 – – The heater or the connected cooling unit receives the control 

signal via the serial interface. Self-tuning is not possible. 

Eprog - 

The heater or the connected cooling unit receives the control 

signal via the E-Prog input. Self-tuning is not possible. 
- Setting requires electronic module. 

 

 
 
 

Important:

  

Set the parameter >Eprog< and the input variables also in the menu 
>Inputs/Outputs< (see page 74). 
 

 

Note:

  

The E-Prog input can only be used either under menu item >Setpoint 
ext< or under menu item >

 

ActVar <. 

 
 

 

Warning:

  

 

The working temperature range of the circulator is determined during configuration. If set 
to >Control<, this range cannot be exceeded. 
If set to >

 

Serial < and >

 

Eprog

 

<, heat or cold is applied to the bath without control. The 

permissible maximum temperature can be exceeded. The user has to take adequate 
precautions for temperature control. 
Materials, such as gaskets or insulations for example, may be damaged or destroyed, if 
the permissible maximum temperature is exceeded. 

The safety and warning functions >

of the instrument must always be used to 

their fullest capacity. 

 

 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание 9352751.N.CSA/UL

Страница 1: ...09 07 18 Operating Manual English Ultra Low Refrigerated Circulator FP51 SL 1 951 2823 V7 06 18 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...ts of ISO 9001 and ISO 14001 Certificate Registration No 01 100044846 Unpacking and inspecting Unpack the circulator and accessories and inspect them for possible transport damage Damage should be rep...

Страница 3: ...26 6 2 Switching on Selecting the language 26 6 2 1 Switching on the circulator 26 6 2 2 Switching on the cooling machine 27 6 3 Start Stop 27 7 Setting of temperatures 28 8 Safety installations warni...

Страница 4: ...65 9 8 1 Limits for internal control and external control 66 9 8 2 Limits for external control 67 9 9 Analog inputs outputs 68 9 9 1 Control connector 72 9 9 2 Outputs of the connector REG E PROG 73...

Страница 5: ...ol to the specific application Absolute Temperature Calibration ATC3 provides high temperature stability in the bath With the 3 point calibration an offset is adjusted at three temperatures to ensure...

Страница 6: ...instructions for the operator You have received a product designed for industrial use Nevertheless avoid strikes to the housing vibrations damage to the operating element panel keypad display and con...

Страница 7: ...a Attention Observe instructions operating manual safety data sheet 2 or Mandatory label M018 Colors blue white Carefully read the user information prior to beginning operation Scope EU 2 Semi S1 0701...

Страница 8: ...trash can to be disposed of separately in an environmentally friendly manner Contact an authorized waste management company in your country Disposal with household waste unsorted waste or similar coll...

Страница 9: ...FP51 SL 9 2 3 EC Conformity www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 10: ...Operator responsibility Safety recommendations 10 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 11: ...rature selection digital via keypad indication on LCD DIALOG DISPLAY C F remote control via PC indication on monitor Temperature indication VFD COMFORT DISPLAY C F Resolution C 0 01 ATC3 INT EXT C 3 9...

Страница 12: ...tions Early warning system for low level float switch High temperature warning function optical audible in intervals Low temperature warning function optical audible in intervals Supervision of workin...

Страница 13: ...it regulates the emission limits use and recovery of these substances It also contains requirements for operators of systems which require contain these substances to function Under Regulation 517 20...

Страница 14: ...hese instructions are not followed serious injury and danger to life could result Caution Describes a possibly dangerous situation If this is not avoided slight or minor injuries could result A warnin...

Страница 15: ...while it is hot Check the temperature of the bath fluid prior to draining e g by switching the unit on for a short moment x Use suitable connecting tubing x Avoid sharp bends in the tubing and mainta...

Страница 16: ...aces in operation moving machine parts etc x Do dangerous steams or gases arise when heating Is an exhaust needed when working x What to do when a dangerous substance was spilled on or in the unit Bef...

Страница 17: ...rol mode in the menu Control internal or external Line 2 Working temp setpoint constantly S xxx xx Line 3 Actual value E external or I internal Alternating with the display in line 1 Use the keys to i...

Страница 18: ...es adjustable via the button in the menu PUMP 3 LCD DIALOG DISPLAY Line 1 Setpoint and origin of setpoint programming Key RS232 or RS485 ext Pt100 EProg Line 2 Actual value internal or external identi...

Страница 19: ...Correction function for parameters or values prior to OK immediately back to standard display icon for keep key pressed down 4 3 1 Key Up Down temperature increase decrease setpoint Push key quickly f...

Страница 20: ...nd temperature recorder output 13 Connector for supplementary pump or solenoid valve 230 V max 1 25 A No control voltage in the OFF condition 14 Connector solenoid valve 1 No control voltage in the OF...

Страница 21: ...9 kg of refrigerant R507 1 m3 of space is required x Do not set up the unit in the immediate vicinity of heat sources and do not expose to sun light x Before operating the unit after transport wait ab...

Страница 22: ...0 0 220 Flash point C 190 230 Fire point C 216 274 Color clear light brown See website for list of recommended bath fluids ATTENTION The maximum permissible viscosity is 70 mm2 s Caution Fire or other...

Страница 23: ...r a temperature coil and fasten with tube clamps to prevent slipping Temperature application to external open systems S Suction pump connection D Pressure pump connection K Float H Height adjustment T...

Страница 24: ...tubing 8 mm inner dia 35 C to 200 C 8930112 1 m Viton tubing 12 mm inner dia 35 C to 200 C 8930410 1 m Insulation for tubing 8 mm inner dia 50 C to 100 C 8930412 1 m Insulation for tubing 12 mm inner...

Страница 25: ...re that no liquid enters the interior of the circulator x Recommended maximum filling level with water as bath fluid 30 mm below the tank rim x Recommended maximum filling level with bath oils 40 mm b...

Страница 26: ...on of the built in fan For proper cooling of the compressor the air must be drawn in from the front of the unit x Connect the circulator with mains power cable 17a to the mains outlet 18 x Connect the...

Страница 27: ...ched off if the current working temperature is increased by 30 C cooling is not required the heater operates at full power 800 W for longer than 5 minutes x It is switched on if cooling is necessary t...

Страница 28: ...er control indicator flashes Setpoint3 Change the value from 70 00 C to 80 00 C See standard display line 1 x Use the numeric keypad to enter and Then press enter to store the value Notice Refer to SE...

Страница 29: ...he parameter is to be retained Possible parameters warning or alarm OverTemp SubTemp x Set value on numeric keypad and confirm by pressing Example 83 00 C Press the keys and and confirm by pressing x...

Страница 30: ...application task adherence to a working temperature value Setpoint is to be supervised then set high and low temperature warning values In the example below the Setpoint of 85 C is surrounded by the...

Страница 31: ...mp The LCD DISPLAY shows the message x Setting ALARM Temperature limit with shutdown of heater and circulating pump An audible alarm sounds continuous tone and a message appears on the VFD COMFORT DIS...

Страница 32: ...cording to IEC 61010 2 010 is triggered a complete shutdown of the heater and circulating pump is effected A continuous alarm tone sounds and a message ALARM CODE 01 appears on the VFD COMFORT DISPLAY...

Страница 33: ...RS232 Setpoint ext Switch over external setpoint setting Autostart AUTOSTART on off Off Mode Pump on off ACTVAR Switch over input variable Time Date setting time and date Language english deutsch Unit...

Страница 34: ...of temperature and capacity Page 66 Setpoint Max Min Maximum and minimum setpoint Heating Max Set maximum heating Cooling Max Set maximum cooling Internal Max Min Limitation of the temperature range U...

Страница 35: ...ct the Step at which the program should be started 0 59 Runs Select the number of program repetitions 1 99 Program End Status at program end StdBy PSetp Setp 1 Setp 2 Setp 3 RUN Start time now time L...

Страница 36: ...confirm by pressing Example hh mm x Set day and month using the numeric keypad and confirm by pressing Example TT MM x Set year using the numeric keypad and confirm by pressing Example JJJJ Confirm S...

Страница 37: ...ample setpoint 2 58 00 C The circulator operates and controls to this temperature at the end of the profile until the key is pressed or a new program is started 9 1 2 Indication after successful start...

Страница 38: ...t and time period set for the corresponding section are thus stopped at the values presently achieved The instrument is put on hold and the message pause flashes on the LCD DIALOG DISPLAY LA program c...

Страница 39: ...If the starting time is not yet reached the programmer is put on hold again If the starting time is exceeded the programmer starts immediately 9 1 6 Termination of a program LA program can be termina...

Страница 40: ...ent indicates that a number needs to be entered Under the submenu Edit enter a program number Six programs may be stored numbers 0 to 5 When the program is running only sections with complete informat...

Страница 41: ...sing and Level 3 L For the selected section ST a setpoint SP and the duration TI or the gradient GRD are entered in the right half of the input window or Input window for ST SP and TI GRD 1 Set the de...

Страница 42: ...red PrgNr with and Example PrgNr 0 Level 3 Input window delete sections x Select menu items from Step and to Step one after the other by pressing and x Select the desired section for every menu item b...

Страница 43: ...K Level 2 Submenu Print x Set menu item Print PrgNr by pressing and x Set desired PrgNr by pressing and Example 4 Level 3 x Confirm Print by pressing x Confirm Print by pressing L Printing and end of...

Страница 44: ...display for pump pressure LIndication of the adjusted pump pressure stage in the OFF mode LIndication of the effective pump pressure stage rotational speed after start To protect the pump motor the ro...

Страница 45: ...ch Remote to on L Control display in the top line for Remote L Connect SERIAL with PC x Switch over setpoint setting by pressing and External setpoint setting Possible parameters off Eprog PT100 x Swi...

Страница 46: ...serial L Set desired parameters with and L Programming of variables for the parameters serial or EPROG is only accepted if the unit is in Start mode Time Date Year see next page Language Possible para...

Страница 47: ...ssible parameters Time hour min Date Day Mon Year x Succesively set time date and year using the numeric keypad and confirm by pressing Example Only the actual time is displayed The internal real time...

Страница 48: ...ddition to the serial interface via remote control the circulator offers the possibility to adjust the setpoint via analog interface ext Pt100 or REG E PROG Possible parameters off Setpoint is set via...

Страница 49: ...he interface Should such a safety standard not be required the AUTOSTART function automatic start mode may be activated thus allowing the unit to be started directly by pressing the mains power switch...

Страница 50: ...ot possible Eprog The heater or the connected cooling unit receives the control signal via the E Prog input Self tuning is not possible Setting requires electronic module Important Set the parameter E...

Страница 51: ...lish Possible parameters deutsch english 9 4 8 Unit Temperature values can be displayed in C or F With this change all adjusted temperature values are converted and displayed in the new unit Possible...

Страница 52: ...Press the key if a parameter is to be retained prior to OK Level 2 Parameter Level or x Switch over control type by pressing and x Possible parameters intern extern L The control type can be adjusted...

Страница 53: ...l x The value flashes Set by using the numeric keypad and confirm by pressing Example Setting range 0 0 5 0 x The value flashes Set by using the numeric keypad and confirm by pressing Setting range 0...

Страница 54: ...the setpoint to be set via the ext Pt100 socket Notice Place the external sensor into the temperature controlled medium and securely fix the sensor Pt100 M R Accessory Pt100 external sensor Order No...

Страница 55: ...g only when the temperature application system changes permanently 9 5 3 Dynamik intern This parameter affects the temperature sequence only in case of internal control C t aperiodic normal setpoint t...

Страница 56: ...ually set via the keypad if necessary to allow optimal control performance Setting range internal external 0 1 99 9 K Proportional range Xp The proportional range is the temperature range below the se...

Страница 57: ...t up curve reveals inappropriate control settings Inappropriate settings may produce the following heat up curves Xp too low C t Tv Tn too low C t Xp too high or Tv too high C t Tv Tn too high or Xp t...

Страница 58: ...ior to OK x The parameter flashes switch by pressing and Possible parameters RS232 RS485 Address RS485 0 to 127 x The parameter flashes switch by pressing and Possible parameters 4800 9600 19200 38400...

Страница 59: ...the bath temperature at the location of the temperature sensor CT is determined at the respective adjusted working temperature This value is then set on the circulator in the menu ATCalibration under...

Страница 60: ...x The parameter flashes switch by pressing and L A 1 point 2 point or 3 point calibration can be carried out The value TmpValue is only indicated L In addition the measured temperature value CalValue...

Страница 61: ...rm by pressing L If only a 2 point calibration is carried out the following menu items are not indicated anymore The value TmpValue is only indicated L In addition the measured temperature value CalVa...

Страница 62: ...ent working temperature also the one set via interface 9 7 3 Type 1 2 3 point A 1 point 2 point or 3 point calibration can be carried out First geometrically define the location for calibration measur...

Страница 63: ...Value X by using the keypad CalValue 1 79 70 C CalValue 2 119 50 C CalValue 3 159 30 C 4 Calibration L Press the key if a parameter is to be retained prior to OK L Setting is required only for the fir...

Страница 64: ...the numeric keypad and confirm by pressing Return to 2 Set working temperature value 160 00 C The value TmpValue is only indicated L In addition the measured temperature value CalValue X is saved dur...

Страница 65: ...eter flashes Set by using the numeric keypad and confirm by pressing x The parameter flashes Set by using the numeric keypad and confirm by pressing x The parameter flashes Set by using the numeric ke...

Страница 66: ...temperature range effects the temperature setting in the menu with the key Only setting of working temperatures which lie within the determined limits is possible Existing settings for Setpoint 1 2 3...

Страница 67: ...fluid at low temperatures For refrigerated circulators Freezing protection when using water as bath fluid Upper band and Lower band The band limitation is active during external control Varied practic...

Страница 68: ...enoid valve SV SV Output Control connector 14 230 V max 1 25 A Channel 1 voltage output for recorder V Channel 2 voltage output for recorder V Channel 3 current output for recorder mA EPROG external p...

Страница 69: ...ing x The value flashes Set by using the numeric keypad and confirm by pressing Define the output value for channel 2 Possible parameters ActInt ActExt Power Setpoint x The parameter flashes switch by...

Страница 70: ...e value flashes Set by using the numeric keypad and confirm by pressing x The value flashes Set by using the numeric keypad and confirm by pressing Possible parameters 0 20 mA 4 20 mA x The parameter...

Страница 71: ...ing x The value flashes Set by using the numeric keypad and confirm by pressing L The value setpoint is only indicated Parameter Level Possible parameters inactive active x The parameter flashes switc...

Страница 72: ...be adapted to solenoid valves for different tasks Backflow In the OFF condition the control output 14 is not powered See Backflow safety device on page 24 C Pulse The control output 14 can be used to...

Страница 73: ...nd 10 V to the highest necessary temperature or power rating required as an output value C Current output channel 3 Assign the current values 0 mA or 4 mA to the lowest and 20 mA to the highest temper...

Страница 74: ...mmer E PROG input of the circulator can be matched to the output signal of the external voltage or current source Voltage voltage input Current current input L Value Setting the LOW value See below Fi...

Страница 75: ...te keys on the keypad of the instrument Example A 200 qC and set by pressing LExample B in the diagram illustrates that the end point values are freely selectable e g 8 mA and 16 mA Example out of dia...

Страница 76: ...for using the STAND BY input The standby function can be used in conjunction with the AUTOSTART feature 1 If the Autostart function is NOT turned ON the standby input is used as described above 2 If...

Страница 77: ...gnments Meaning of the terms under menu item Function The circulator is in condition StandBy or Alarm For Type normal pins 2 and 3 are connected in any case according to the selected Function For Type...

Страница 78: ...the liquid level is insufficient Replenish the bath tank with the bath fluid Tube breakage has occurred insufficient filling level due to excessive bath fluid pumped out Replace the tubing and repleni...

Страница 79: ...ling water The first stage compressor does not work After a short cooling interval the compressor motor will be automatically reconnected and the message CODE 21 will no longer appear The second stage...

Страница 80: ...e configuration Then contact an authorized JULABO service center Disturbances that are not indicated The electronic pump motor is protected against overload by an electronic current limiter If the vis...

Страница 81: ...cable is electrically connected to the plug housing and the sensor tube SERIAL RS232 RS485 serial interface This port can be used to connect a computer with an RS232 or RS485 cable for remote control...

Страница 82: ...oller for cooling water Pin assignment Pin Signal only with attached JULABO equipment 1 24 V I max current 25 mA 2 0 V 3 Alarm relay 4 Reserved do not use 5 Cooling pulse ALARM Alarm output for extern...

Страница 83: ...V 0 to 20 mA 5 Current output Channel 3 0 to 20 mA 4 to 20 mA 6 GND for Progammer 0 V Control connectors 13 14 The control output 13 is not powered in the OFF condition An external pump motor can be...

Страница 84: ...SPLAY will read R OFF REMOTE STOP The circulator is now operated via the computer In general the computer master sends commands to the circulator slave The circulator sends data including error messag...

Страница 85: ...5 LF Notice OUT commands Temperature values can be displayed on the circulator in C or F The settings sent via the interface must also be made in the selected unit according to this setting 12 3 List...

Страница 86: ...l 0 5 0 OUT_PAR_06 xxx Xp control parameter of the internal controller 0 1 99 9 OUT_PAR_07 xxx Tn control parameter of the internal controller 3 9999 OUT_PAR_08 xxx Tv control parameter of the interna...

Страница 87: ...Pump stage corresponding to the effective rotational speed after start IN_SP_08 none Value of a flow rate sensor connected to the E Prog input IN_SP_09 none Value of a pressure sensor connected to th...

Страница 88: ...0 Selftuning off 1 Selftuning once 2 Selftuning always IN_MODE_03 none Type of external programmer input 0 Voltage 0 V to 10 V 1 Current 0 mA to 20 mA IN_MODE_04 none Internal external temperature co...

Страница 89: ...ternal error when reading or writing the I2 C bus 08 INVALID COMMAND Invalid command 09 COMMAND NOT ALLOWED IN CURRENT OPERATING MODE Invalid command in current operating mode 10 VALUE TOO SMALL Enter...

Страница 90: ...setpoint programming set to external Pt100 40 NIVEAU LEVEL WARNUNG Low liquid level warning in the internal reservoir 13 Installation of electronic module with analog connectors Caution Always turn of...

Страница 91: ...ALARM x Connect the circulator to the computer using an interface cable x Start the EasyBlackBox program The program asks for the port used COM1 and the baud rate of the unit You do not have this info...

Страница 92: ...tside of the unit using a wet cloth and low surface tension water The circulator is designed for continuous operation under normal conditions Periodic maintenance is not required The tank should be fi...

Отзывы: