Operating procedures
26
6. Operating
procedures
6.1. Power
connection
Caution:
x
Only connect the unit to a power socket with earthing contact (PE – protective
earth)!
x
The power supply plug serves as safe disconnecting device from the line and must
be always easily accessible.
x
Never operate equipment with damaged mains power cables.
x
Regularly check the mains power cables for material defects (e.g. for cracks).
x
We disclaim all liability for damage caused by incorrect line voltages!
Check to make sure that the line voltage matches the supply voltage
specified on the identification plate.
Mains power connection:
3x400V/N/PE
3x230V/PE
L
Note the direction of rotation of the built in fan!
For proper cooling of the compressor, the air must be drawn in
from the front of the unit.
x
Connect the circulator with mains power cable (17a) to the mains
outlet (18).
x
Connect the control cable (9) to the
7
sockets (9a, 9b).
x
Connect the refrigerated circulator with mains power cable (17b) to
the mains socket.
6.2. Switching on / Selecting the language
6.2.1. Switching
on
the
circulator
Switching on:
x
Turn on the mains power switch (1a).
L
The unit performs a self-test.
Then the software version (example: V 1.xx-xxxx) appears. The
display „
OFF
“ or „
R
OFF
“ indicates the unit is ready to operate.
L
The circulator enters the operating mode activated before switching
the circulator off:
keypad control mode
(manual operation) or
remote control mode
(operation via personal computer).