FP51-SL
63
Î
1. Set circulator to internal control:
L
The type of control can be set only in the –OFF- mode.
80.00 °C
120.00 °C
160.00 °C
2.
Set working temperature setpoint
See standard display line 1
x
Example 80.00 °C
Press the keys
and
and confirm by pressing
.
x
The bath is heated up.
Wait for approx. 5 minutes until the temperature is constant.
3. Reading of temperature measuring device
Read the value of measuring point CT on the device and enter under menu
item > CalValue X< by using the keypad.
CalValue 1< (79.70 °C)
CalValue 2< (119.50 °C)
CalValue 3< (159.30 °C)
Ð
4. Calibration
Ð
L
Press the
key if a parameter is to be retained. (prior to OK)
L
Setting is required only for the first calibration point.
Level 2
Parameter-Level
Î
Set sensor: >intern<.
x
The parameter flashes, switch by
pressing
and
Î
Set to >no<
x
The parameter flashes, switch by
pressing
and
Î
A >3-point< calibration is carried out.
x
The parameter flashes, switch by
pressing
and
Î
The value >TmpValue< is only indicated.
L
In addition the measured temperature
value >CalValue X< is saved during the
next step.
www.
GlobalTestSupply
.com
Find Quality Products Online at: