FP51-SL
93
15. Cleaning / repairing the unit
Caution:
x
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
x
Prevent humidity from entering into the circulator.
x
Electrical connections and any other work must be performed by qualified
personnel only.
In dusty environments, the refrigerating machine’s condenser
should be cleaned by vacuuming every 1 to 2 months. The
condenser is located behind the venting grid on the front of the unit,
which can be taken off by removing the screws.
x
Switch off the unit, disconnect mains power cable.
x
Remove the venting grid.
x
Clean the ribbed condenser with a vacuum cleaner.
x
Replace the venting grid.
x
Switch on the unit.
Cleaning:
For cleaning the bath tank and the immersed parts of the circulator, use low surface
tension water (e.g., soap suds).
Clean the outside of the unit using a wet cloth and low surface tension water.
The circulator is designed for continuous operation under normal conditions. Periodic
maintenance is not required.
The tank should be filled only with a bath fluid recommended by JULABO. To avoid
contamination, it is essential to change the bath fluid from time to time.
Repairs
Before asking for a service technician or returning a JULABO instrument for
repair, please contact an authorized JULABO service station.
When returning the unit:
x
Clean the unit in order to avoid any harm to the service personnel.
x
Attach a short fault description.
x
During transport the unit has to stand upright. Mark the packing correspondingly.
x
When returning a unit, take care of careful and adequate packing.
x
JULABO is not responsible for damages that might occur from insufficient
packing.
JULABO reserves the right to carry out technical modifications with repairs for
providing improved performance of a unit.