Joerns Support Surface
Arise
®
LAL Mattress Replacement,
Arise LAL Overlay, Arise 1000, Arise 1000EX
15
Wash hands often and well, including after
removal of gloves.
Patient care equipment that is used in isolation
areas should be disinfected in accordance with
all internal policies and procedures regarding
such equipment.
Disinfecting
When there is visible soilage, and between patients,
it is recommend that the unit and mattress be
disinfected with a tuberculocidal disinfectant.
Disinfectant should be registered with the
Environmental Protection Agency (EPA).
Use rubber gloves and eye protection.
Prepare detergent/disinfectant (registered by
EPA as hospital disinfectant) solution according
to instructions on label for correct use-dilution.
With support surface deflated, thoroughly wipe
down entire mattress, as air cells will lie flat.
Be sure to reach all areas underneath and in-
between air cells. Allow to air dry.
If dust or other soiling has accumulated along
air hoses, remove using swabs moistened with
detergent/disinfectant as necessary. Allow all
components to air dry. Wrap mattress in plastic
and return to storage area.
Thoroughly wipe down outside of control unit
and allow to air dry. Cover with plastic and return
to storage area
Remove gloves and dispose; wash hands.
Therapy Pad
The therapy pad can be wiped down with a
disinfectant solution or a mild detergent with a damp
cloth. If heavily soiled, the therapy pad can be
laundered in a washer and dryer with warm water
(no more than 120º Fahrenheit). A
non-bleach detergent should be used sparingly.
Wipe dry or allow to air dry.
Steam Cleaning
Do not use any steam cleaning device on the unit.
Excessive moisture can damage mechanisms in
this unit.
4.
5.
1.
2.
3.
4.
5.
6.
Filter Cleaning
Check the air filter on the rear of the unit regularly
for buildup of dust/dirt. If buildup is visible turn off
the control unit and disconnect the power cord
from the wall outlet. Remove the filter by grasping
the filter pulling outward. Replace with the second
supplied filter. Ensure the replaced filter covers the
entire filter region.
Hand-wash the removed filter in warm soapy water
and allow to air dry. When dry, store the filter in a
safe place for the next filter maintenance.
Maintenance
Warning:
Only facility-authorized personnel
trained by Joerns Healthcare should perform
preventative maintenance. Preventative
maintenance performed by unauthorized
personnel could result in personal injury or
equipment damage.
Any maintenance done without Joerns’
authorization will invalidate any warranties on this
product.
Storage and Care
When the product is not in use, properly store
the power cord. Failure to do so could result in
personal injury.
Note:
Clean the Arise
®
as described in the
previous section prior to storage.
Control Unit
The power cord may be wrapped around the unit
for convenience. Wrap the unit in a plastic bag
for dust resistance then store the unit in an area
appropriate for an electronic medical device.
Support Surface
Gently roll up the support surface, expelling any
residual air, for temporary storage. The mattress
should be wrapped in plastic and/or a clean bag
for storage.