Joerns Support Surface

Arise

®

 LAL Mattress Replacement,  

Arise LAL Overlay, Arise 1000, Arise 1000EX

12

Additional Features

Warning:

 A possible fire hazard exists. 

This product is suitable for use with oxygen 
administering equipment of the nasal, mask, or 
half bed-length, tent-type only. To prevent personal 
injury or equipment damage, ensure that the 
oxygen tent does not extend below the mattress.

CPR

The hose connection at the control unit is marked 

CPR

. Disconnect the hose from the control unit. 

CPR connection style may vary by model. To  
re-inflate the air cells, reattach the hose to the 
control unit.

Transport

To transport the patient in bed, turn the control unit 
off. Unplug the power cord from the outlet. Do not 
disconnect the hose connection at the control unit. 
The lower support cell will prevent the patient from 
bottoming out for up to 12 hours. Some models 
incorporate a foam safety cell.

For the Arise

®

 1000 EX mattress, the control unit 

can be disconnected from the mattress as the 3" 
foam base will continue to support the patient. If 
the Bari10A bed frame is to be reduced in width 
for transport it may be necessary to deflate the 
CairRails. This can be done at the mattress control 
panel (Figure 2, pg 10) by turning the lower right 
control knob down.

Power Failure

The Arise control unit has an advanced power 
failure notification alarm and Comfort Adjust 
setting memory capability. Upon power failure 
the control unit will flash a red Alarm light on the 
keypad and sound an audible signal every two (2) 
seconds for up to twenty (20) minutes. The power 
interruption feature keeps the current settings in 
memory based on the frequency of use and level 
of battery charge. The Alarm Reset will turn off 
audible and visual alarm.

Power Interruption (Brown out)

In the event of a brief power interruption, up to 
twenty (20) minutes, the Arise control unit will 
automatically restart and reset to the previous 
Comfort Adjust settings. The power interruption 
feature keeps the current settings in memory based 
on the frequency of use and level of battery charge.

Power Loss

In the event of a power failure in excess of twenty 
(20) minutes, the control unit’s stand by light will be 
illuminated after power restoration. Press the Power 
button and reset the preferred mode and comfort 
level as described in the “Operation” section. 

Safety cell

In the event of a power failure, the lower support 
cells should remain inflated for up to 12 hours as 
long as the control unit remains connected. In case 
of an extended power failure, transfer the patient to 
a hospital mattress or other surface.

CairRails

Integrated CairRails risk management 
air  bolsters offer a bilateral side bolster solution 
designed to address healthcare’s growing concerns 
of liability in relation to patient falls and entrapment. 
CairRails are being recognized by some of the 
nations leading healthcare systems for improving 
their patient safety and risk management programs.

CairRails are recommended for patients requiring 
additional support during patient care and transfer. 
CairRails can help reduce costs while ensuring 
optimal clinical outcomes and increasing  
patient safety.

Note: 

When inflating CairRails, it is recommended 

that the control unit be in 

Autofirm

 mode to achieve 

optimal results.

Electronic Copy: Reference Only

Содержание arise1000

Страница 1: ...r Service Manual Joerns Support Surface Arise LAL Mattress Replacement Arise LAL Overlay Arise 1000 Arise 1000EX transforming wound management To avoid injury read user s manual before using Electroni...

Страница 2: ...stem always ensure that the patient is positioned properly within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Dan...

Страница 3: ...aps and spaces between bed system components and clinical guidance for assessment and implementation of bed side rails in various health care settings Entrapment zones involve the relationship of bed...

Страница 4: ...and Monitoring Additional Features 12 13 CPR Transport Power Failure CairRails Optional Accessories Troubleshooting 13 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Con...

Страница 5: ...from these damaging forces 5 Introduction The Arise provided by Joerns Healthcare is a unique therapy system that provides pressure relief by combining low air loss with pulsation Low air loss therap...

Страница 6: ...l times The use of the Arise for these patients should be considered on an individual basis and discussed with the attending physician Features The Arise is comprised of two components Therapy control...

Страница 7: ...ut for most patients within weight limit Integrated low pressure heel section provides lower interface pressures Durable base tub is constructed from 100 heavy weight 1680 denier nylon with a 1 5 oz u...

Страница 8: ...proper ground at the outlet Inadequate grounding at electrical outlets can occur even if there is a ground wire Wires can become loose over time at the connection to the outlet Note To install new wir...

Страница 9: ...lls facing up Secure the straps around the head and foot ends of the mattress Fasten the two straps by the seat section securely to the sleep deck of the bed frame 4 5 If the therapy pad is not alread...

Страница 10: ...flated as required Locate the control knob Figure 2 on the lower right of the control panel on the mattress Next inflate deflate the CairRails by moving the control knob to the up inflate or down defl...

Страница 11: ...urn to the mode it was in prior to Autofirm either therapy or pulsate in approximately 12 minutes The Arise has two low air loss therapy modes Therapy and Pulsate Therapy The unit starts in the Therap...

Страница 12: ...to twenty 20 minutes The power interruption feature keeps the current settings in memory based on the frequency of use and level of battery charge The Alarm Reset will turn off audible and visual alar...

Страница 13: ...ke sure the Lockout function is disabled To disable press the up and down comfort adjust arrows simultaneously 1 2 3 4 Nursing Procedures Recommended Linen Special linens are not necessary for the Ari...

Страница 14: ...s Early intervention may be essential to preventing serious skin breakdown Bed Height To minimize the risks of falls or injury the patient surface should always be in the lowest practical position whe...

Страница 15: ...warm water no more than 120 Fahrenheit A non bleach detergent should be used sparingly Wipe dry or allow to air dry Steam Cleaning Do not use any steam cleaning device on the unit Excessive moisture...

Страница 16: ...8 122 cm W 80 203 cm H 8 5 22 cm D Arise 1000EX Mattress 36 42 48 91 107 122 cm W 80 88 203 224 cm L 8 5 22cm D Electrical Specifications USA 120V AC 60 Hz 4A less than 1A running Environmental Condit...

Страница 17: ...www joerns com info joerns com Joerns Support Surface Arise LAL Mattress Replacement Arise LAL Overlay Arise 1000 Arise 1000EX 17 Notes Electronic Copy Reference Only...

Страница 18: ...www joerns com info joerns com Joerns Support Surface Arise LAL Mattress Replacement Arise LAL Overlay Arise 1000 Arise 1000EX 18 Notes Electronic Copy Reference Only...

Страница 19: ...www joerns com info joerns com Joerns Support Surface Arise LAL Mattress Replacement Arise LAL Overlay Arise 1000 Arise 1000EX 19 Notes Electronic Copy Reference Only...

Страница 20: ...authorized in writing by Joerns Healthcare will void this warranty Damage caused by use in unsuitable environmental conditions abuse or failure to maintain the product in accordance with user and serv...

Отзывы: