PAGE 111 of 177
CX0344 Rev 05.3
1.
Attach the wedges to the probe holders that are to be used
(see “Probe Holder
Setup” on page 69 for additional details)
.
2.
Affix the probe holders (with attached wedges) to the low profile probe holder
frame. On the frame bar, place the
2
secondary probe holders at the front
and the
1
primary probe holders at the rear
(Fig. 260)
.
TIP:
Due to their larger size, scan results are generally improved when pulling or
dragging phased array wedges.
3.
Mount the low profile probe holder frame to the crawler
(see “Swivel Mount” on
page 65 for additional details)
. When mounting the low profile frame, ensure
the attachment knobs
(Fig. 260)
are at the front (non crawler side).
4.
Release the two swivel mount levers
(Fig. 261)
to position the swivel mount
parallel to the scan surface
(Fig. 262)
. When alignment with scan surface is
achieved, lock the crawler swivel mount levers.
2
1
Fig. 260 - Position primary and secondary probe holders
Fig. 261 - Align swivel mount with scan surface