16
Fig. 1: Registered design no. 27650
h)
Caution
As soon as the device is switched on, the spark discharge creating Plasma is ignited between
the tip of the device and patient´s grounded skin and the device starts a continual control of
the current running through patient. If the current intensity exceeds 1mA due to a hardware
failure or malfunction, the message “SrvC” is shown on the display, the device is switched off
automatically and it cannot be switched on any more. It must be sent to an authorized service
for maintenance.
i)
Cosmetic treatment (Lower intensities 1-5)
Approach the tip of the device (1) to a treated place protected by an even layer of conductive
indifferent gel or by a layer of a conductive serum (for example hyaluronic acid) as near as 2
mm above the gel. This causes a dielectric breakdown of the air between the tip of the device
and the gel protecting patient´s skin and a flow of spark discharges is generated. These sparks
treat patient´s skin. The time of treatment should not exceed 30 minutes. During the treatment,
a bar graph shows the effectivity of the discharge, i.e. to what extend the possible effect of the
discharge is used.
j)
Medical treatment (desiccation, fulguration, coagulation) at the maximum intensity
level of 6
–
8.