14
6)
A dash before and after the number indication the intensity of spark
discharge
7)
ON/OFF button
8)
Slot for connection of grounding cable for the connection of patient
9)
Power Indicator
10)
Slot for power supply cable
11)
Button for decrease of discharge intensity
12)
Button for increase of discharge intensity
Before every start of the device, carefully check the power cable, connecting cables,
disposable grounding electrode with its connecting cable and device including its tip. If
you encounter any damages, do not switch the device on and call the authorized service!
a)
Plug the adaptor the power source (230V, 50Hz) and plug the other side of the supply
connector into the slot of Jett Plasma Lift Medical (10). The device switches on automatically
and a mess
age “TAPE”
appears on the screen, telling you that the patient must be grounded.
You can also switch the device off and on by pushing the ON/OFF button (7) for 2-3 seconds.
b)
When the patient is ready for the treatment, stick a disposable grounding electrode on his
or her right shoulder blade or on his right hand. Plug the connector of the grounding cable into
the grounding slot (8) and connect its other side to the disposable grounding electrode. The
message “TAPE”
disappears from the display and the lowest inten
sity level „
-1-
„ appears.
c)
Setting the spark discharge intensity
After switching the device on, default intensit
y setting is 1. By pushing the “+”
button, you
can increase the intensity. Between the two dashes, numbers
“2, 3, 4, 5”
appear one by one.
These are safe intensities that do not cause the destruction of skin cells. Decrease of safe
intensit
ies can be done by pushing the “
-
“
button (11). As soon as you set the first possibly
dangerous intensity level “
-6-
“ (able t
o destroy cells), the device sets the intensity level
automatically to “
-1-
“
and the dashes before and after start to blink. Now you have 6 seconds
to set the intensity level to 6 or higher.
d)
Discharge ignition
If the intensity of the spark discharge is set already, it is possible to set a higher voltage (6kV)
at the tip of the device by pushing the button Play/Pause (3). As soon as the tip approaches the
patient´s grounded skin to a distance below 2 mm, the spark discharge ignites.
e)
Turning the discharge off
If you want to turn the high voltage at the tip of the device off, you must push the button
Play/Pause (3). The change of the intensity of the discharge can be made only when the
device is switched off, because there isn´t the 6 kV voltage at the tip.
f)
Quicker changing of dangerous intensities