23
13.2 Warranty
This product is covered by the warranty including the material imperfections and production
faults for one year after the supply of the product.
The manufacturer agrees to repair or change the product or its faulty parts in the time
covered by the warranty, it these repairs are done by the manufacturer or its authorized agents.
The manufacturer takes responsibility for the security, reliability and effectivity of the
product only if all repairs, changes of parts of alterations of the product are made by the
authorized staff, if the device was used in accordance with this manual, if the parameters of
the electricity in the room where the device is plugged fit all the requirements set by law.
If the product is sent to the manufacturer, the transport is paid by the user.
13.3 Storage, maintenance and safety technical checks
How to store and maintain the product after use:
- Storage temperature: 5 -
40 °C
- Store indoors in well-aired places and protect from direct sunlight and humidity.
- Store unplugged from power supply
Regular maintenance, cleaning a periodical safety technical checks in given time.
-
Maintenance
–
the operators are obliged to follow all the instructions in this manual.
The integrity of the device cannot be violated, maintenance can be done exclusively at
the annual checks by the authorized service.
-
The applicators must be disinfected after each use by common disinfectants stated
below. The body of the device can be disinfected by moist cloth only.
Disinfection:
After each treatment, a thorough disinfection of the used applicator must be done by one of
these disinfectants:
CUTASEPT F
SEKUSEPT EXTRA N
SEKUSEPT PLUS
The applicator must be wiped carefully. It is necessary to alternate the disinfectants!
Safety technical checks
If periodical safety technical checks (PSTC) are given by national laws or norms, they should
be done annually. PSTC can be done only by a service worker o ran authorized service. If the
safety check has not been done by the manufacturer or an authorized service organization in a
set period, the device must not be used until the check is done.
Service, maintenance or PSTC must not be done during the treatment of a patient.