background image

23 

 

13.2 Warranty 

 

This product is covered by the warranty including the material imperfections and production 
faults for one year after the supply of the product.  

 

The manufacturer agrees to repair or change the product or its faulty parts in the time 

covered by the warranty, it these repairs are done by the manufacturer or its authorized agents.   

 

The manufacturer takes responsibility for the security, reliability and effectivity of the 

product only if all repairs, changes of parts of alterations of the product are made by the 
authorized staff, if the device was used in accordance with this manual, if the parameters of 
the electricity in the room where the device is plugged fit all the requirements set by law.  

If the product is sent to the manufacturer, the transport is paid by the user.  

13.3 Storage, maintenance and safety technical checks  

 

How to store and maintain the product after use: 

- Storage temperature: 5 - 

40 °C

 

- Store indoors in well-aired places and protect from direct sunlight and humidity.  
- Store unplugged from power supply  
 

Regular maintenance, cleaning a periodical safety technical checks in given time. 

-

 

Maintenance 

 

the operators are obliged to follow all the instructions in this manual. 

The integrity of the device cannot be violated, maintenance can be done exclusively at 
the annual checks by the authorized service. 

 

-

 

The applicators must be disinfected after each use by common disinfectants stated 
below. The body of the device can be disinfected by moist cloth only.  

 

 

 
Disinfection: 

After each treatment,  a  thorough disinfection of the used applicator must be done by one of 
these disinfectants:  
CUTASEPT F 
SEKUSEPT EXTRA N 
SEKUSEPT PLUS 
 
The applicator must be wiped carefully. It is necessary to alternate the disinfectants! 

Safety technical checks 

If periodical safety technical checks (PSTC) are given by national laws or norms, they should 
be done annually. PSTC can be done only by a service worker o ran authorized service. If the 
safety check has not been done by the manufacturer or an authorized service organization in a 
set period, the device must not be used until the check is done.  

Service, maintenance or PSTC must not be done during the treatment of a patient.  

Содержание Plasma Lift Medical

Страница 1: ...JPL Medical Date 9 9 2015 Jett Plasma Lift Medical User Manual Before you start using Jet Plasma Lift read this manual and make sure you understand all instructions and warnings and follow them close...

Страница 2: ...the Manual 10 6 Warnings and precautions 10 7 Operation of the device 13 8 Definition of plasma 17 8 1 The nature of discharge generating 18 9 Effect on cells 18 10 Use 19 11 Types of treatment 19 12...

Страница 3: ...vice and conductively connected to the device skin of a patient and maintaining the distance of 2mm between the tip and skin this constant distance is provided by a specific applicator a spark dischar...

Страница 4: ...lectrodes solid gel 51 33 mm v 300612979 Flat applicator Golden applicator with pointed tip Cone applicator AC adapter Type DA12 050EU M Input 100 240V 50 60Hz Output 5VDC max 2 0A Power max 12 VA Man...

Страница 5: ...e state The aim of the use are corrections removal of skin structures by the use of sparkle discharges generated by DC current Some of the aforementioned effects indications are not of a clinical char...

Страница 6: ...this is that the energy absorbed by body can be in cases of some applications used for targeted destruction of skin cells At the highest levels 6 7 and 8 this device is used for electro fulguration e...

Страница 7: ...used in classical surgery dermatology gynecology or dental medicine At medium intensities 5 6 when there is no considerable destruction of skin cells the device is used for the treatment of small vein...

Страница 8: ...device it must not be disposed of in communal waste Graphic sign Double insulation Graphic sign Use indoors Graphic sign Protection from a lightning Graphic sign ground galvanic connection with the de...

Страница 9: ...r storage and transport Graphic sign Range of pressure min 50 kPa max 106 kPa Graphic sign Humidity limitations min 30 max 75 Graphic sign Protect from rain Labels at the rear side of the device Flat...

Страница 10: ...explanations before you start to operate the device PRECAUTION introductory precautions All parts of the device must be connected in their places in the designed positions Use of a power adaptor other...

Страница 11: ...stick to their body during the whole time of the treatment If the patient is nt grounded the display shows the warning TAPE and the generator the Plasma is disconnected automatically TErr Two buttons...

Страница 12: ...rounding Caution Before the treatment begins the patient must be grounded by a grounding disposable electrode which is stuck on the right shoulder blade Do not touch the tip Caution Neither operator n...

Страница 13: ...the device can destroy skin cells at the intensity of 6 7 and 8 Before every application make sure twice that you have set the right intensity That means that if you do the treatment of couperosis or...

Страница 14: ...ther side to the disposable grounding electrode The message TAPE disappears from the display and the lowest intensity level 1 appears c Setting the spark discharge intensity After switching the device...

Страница 15: ...ic applicator has the same use but it does not touch skin it touches only the indifferent gel 2 Golden applicator with precise tip It is designed for face treatments and precise removals of skin pertu...

Страница 16: ...ce i Cosmetic treatment Lower intensities 1 5 Approach the tip of the device 1 to a treated place protected by an even layer of conductive indifferent gel or by a layer of a conductive serum for examp...

Страница 17: ...ectrons created by the separation of electrons from the electron shell of gases or by ionization Plasma can be also defined as the fourth form of matter In this case plasma is quasineutral and low tem...

Страница 18: ...ong the membrane and the electric voltage of the membrane is changed The Membrane potential is created and influenced by potassium and sodium cations It is difficult for sodium cations to pass through...

Страница 19: ...ted The spark channel generated by DC voltage is very slim approx 1 mm2 and therefore the treatment and the targeted destruction of cells is very precise The treatment of skin by spark discharge is a...

Страница 20: ...1st and the 2nd treatment there should be 3 free days after every other treatment a longer free period of 1 week is recommended One treatment including preparation time and aftercare takes 1 hour Afte...

Страница 21: ...ow treatment Move the tip of the device firmly slowly and carefully in the area of wrinkles 1 2 mm above the gel as you have been taught at the training If bigger areas are treated use the flat applic...

Страница 22: ...e at the tip to zero after switching off after 1 s Size 245 mm 45 mm Weight approx 350 g Range of relative air humidity 30 75 Range of environment temperature 10 to 40 o C Range of atmospheric pressur...

Страница 23: ...nd humidity Store unplugged from power supply Regular maintenance cleaning a periodical safety technical checks in given time Maintenance the operators are obliged to follow all the instructions in th...

Страница 24: ...of electrical or electronic devices may contain substance that may be harmful for the environment or health Jett Plasma Lift Medical is included in the electric waste and after the end of its lifespa...

Страница 25: ...420 602 766 759 00420 725 385 074 e mail nespor compexbrno cz export compexbrno cz The manufacturer has a system of authorized service centers For more information see www jettcosmetics cz 13 5 The De...

Страница 26: ...or A 0001 15 DATE OF MANUFACTURE 15 2 2015 Compex spol s r o Palack ho t da 105 612 00 Brno Czech Republic tel 00420 602 766 759 www jettmedical cz Golden applicator with a precise tip B 0001 15 DATE...

Страница 27: ...27 C 0001 15 DATE OF MANUFACTURE 15 2 2015...

Отзывы: