background image

 

 

Electrodessication  is  evaporation  of  cell  liquids  with  a  subsequent  destruction  of 

subcutaneous  cells.  Evaporation  is  caused  by  thermal  energy  generated  by  oscillating  spark 
discharge. A relatively low electric power, at maximum 1.8 W is applied and as it affects only 
a small area, it is able to cause the evaporation of cell liquids.  

 

Electrocoagulation is a process in which thermal energy of the spark discharge causes 

evaporation  of  water  in  cells  and  it  disrupts  their  structure.  High  temperature  causes 
coagulation  of  tissue  and  blood  proteins,  which  is  used  for  stopping  a  small  bleeding 

  to 

closure  of  small  veins  with  a  diameter  of  2-3  mm.  The  treated  area  is  stimulated  by  spark 
discharge with a creation of plasma at the distance of 2 mm from the skin. The discharge is 
applied  with  a  smaller  power  (1.8  W),  compared  to  common  electrocauters.  Concerning  the 
fact,  that this  power works  in  a very  thin  discharge beam  and treats  only  a small area, there 
appears a considerably good local coagulation effect without a need to sterilize the application 
tip.  

 

 

Electrodessication,  electrofulguration  or  electrocoagulation  can  be  used  for  the 

treatment of various indications in various medical fields. A detailed summary of application 
possibilities will be a subject of training together with the handover of the device. The device 
can be used in classical surgery, dermatology, gynecology or dental medicine.  

 

At medium intensities (5-6) when there is no considerable destruction of skin cells, the 

device is used for the treatment of small veins, couperosis, scars and stretchmarks or pigment 
spots.  Here,  the  thermal  energy  of  the  spark  discharge  is  used  for  the  removal  of  skin  cells 
with accumulated melanin. These intensities can be satisfactorily used for the veins removal.  

The  spark  discharge  also  opens  Na/K  channels  of  cell  membranes  in  skin  cells. 

Together with this, being generated by AC voltage with a minus pole on the tip, it is able to 
help  with  the  absorption  of  various  serums  of  negative  ions  character  into  the  skin.  Unlike 
other  technologies,  it  works  with  a  higher  energy  and  a  doctor  is  able  to  apply  an  optimal 
amount  of  serum  very  effectively.  The  desired  effect  is  supported  by  the  simultaneous 
opening of Na/K channels and the heat generated by the spark discharge.  

At  the  lowest  intensities  (1-4),  the  spark  discharge  is  used  solely  for  aesthetic 

treatments that can be a supplement to a medical treatment or an extension of the offering of 
medical clinics or ambulances. The spark discharge generates a flow of negative electrons and 
positive ions. The discharge is generated by the DC current and there is a negative pole at the 
tip  of  the  device.  Most  of  cosmetic  preparations  contains  negative  ions  that  are  very  well 
processed  thanks  to  the  polarization.  These  intensities  are  used  for  lifting,  treatment  of 
wrinkles and rejuvenation.  

5. Description of Symbols  

 

Содержание Plasma Lift Medical

Страница 1: ...JPL Medical Date 9 9 2015 Jett Plasma Lift Medical User Manual Before you start using Jet Plasma Lift read this manual and make sure you understand all instructions and warnings and follow them close...

Страница 2: ...the Manual 10 6 Warnings and precautions 10 7 Operation of the device 13 8 Definition of plasma 17 8 1 The nature of discharge generating 18 9 Effect on cells 18 10 Use 19 11 Types of treatment 19 12...

Страница 3: ...vice and conductively connected to the device skin of a patient and maintaining the distance of 2mm between the tip and skin this constant distance is provided by a specific applicator a spark dischar...

Страница 4: ...lectrodes solid gel 51 33 mm v 300612979 Flat applicator Golden applicator with pointed tip Cone applicator AC adapter Type DA12 050EU M Input 100 240V 50 60Hz Output 5VDC max 2 0A Power max 12 VA Man...

Страница 5: ...e state The aim of the use are corrections removal of skin structures by the use of sparkle discharges generated by DC current Some of the aforementioned effects indications are not of a clinical char...

Страница 6: ...this is that the energy absorbed by body can be in cases of some applications used for targeted destruction of skin cells At the highest levels 6 7 and 8 this device is used for electro fulguration e...

Страница 7: ...used in classical surgery dermatology gynecology or dental medicine At medium intensities 5 6 when there is no considerable destruction of skin cells the device is used for the treatment of small vein...

Страница 8: ...device it must not be disposed of in communal waste Graphic sign Double insulation Graphic sign Use indoors Graphic sign Protection from a lightning Graphic sign ground galvanic connection with the de...

Страница 9: ...r storage and transport Graphic sign Range of pressure min 50 kPa max 106 kPa Graphic sign Humidity limitations min 30 max 75 Graphic sign Protect from rain Labels at the rear side of the device Flat...

Страница 10: ...explanations before you start to operate the device PRECAUTION introductory precautions All parts of the device must be connected in their places in the designed positions Use of a power adaptor other...

Страница 11: ...stick to their body during the whole time of the treatment If the patient is nt grounded the display shows the warning TAPE and the generator the Plasma is disconnected automatically TErr Two buttons...

Страница 12: ...rounding Caution Before the treatment begins the patient must be grounded by a grounding disposable electrode which is stuck on the right shoulder blade Do not touch the tip Caution Neither operator n...

Страница 13: ...the device can destroy skin cells at the intensity of 6 7 and 8 Before every application make sure twice that you have set the right intensity That means that if you do the treatment of couperosis or...

Страница 14: ...ther side to the disposable grounding electrode The message TAPE disappears from the display and the lowest intensity level 1 appears c Setting the spark discharge intensity After switching the device...

Страница 15: ...ic applicator has the same use but it does not touch skin it touches only the indifferent gel 2 Golden applicator with precise tip It is designed for face treatments and precise removals of skin pertu...

Страница 16: ...ce i Cosmetic treatment Lower intensities 1 5 Approach the tip of the device 1 to a treated place protected by an even layer of conductive indifferent gel or by a layer of a conductive serum for examp...

Страница 17: ...ectrons created by the separation of electrons from the electron shell of gases or by ionization Plasma can be also defined as the fourth form of matter In this case plasma is quasineutral and low tem...

Страница 18: ...ong the membrane and the electric voltage of the membrane is changed The Membrane potential is created and influenced by potassium and sodium cations It is difficult for sodium cations to pass through...

Страница 19: ...ted The spark channel generated by DC voltage is very slim approx 1 mm2 and therefore the treatment and the targeted destruction of cells is very precise The treatment of skin by spark discharge is a...

Страница 20: ...1st and the 2nd treatment there should be 3 free days after every other treatment a longer free period of 1 week is recommended One treatment including preparation time and aftercare takes 1 hour Afte...

Страница 21: ...ow treatment Move the tip of the device firmly slowly and carefully in the area of wrinkles 1 2 mm above the gel as you have been taught at the training If bigger areas are treated use the flat applic...

Страница 22: ...e at the tip to zero after switching off after 1 s Size 245 mm 45 mm Weight approx 350 g Range of relative air humidity 30 75 Range of environment temperature 10 to 40 o C Range of atmospheric pressur...

Страница 23: ...nd humidity Store unplugged from power supply Regular maintenance cleaning a periodical safety technical checks in given time Maintenance the operators are obliged to follow all the instructions in th...

Страница 24: ...of electrical or electronic devices may contain substance that may be harmful for the environment or health Jett Plasma Lift Medical is included in the electric waste and after the end of its lifespa...

Страница 25: ...420 602 766 759 00420 725 385 074 e mail nespor compexbrno cz export compexbrno cz The manufacturer has a system of authorized service centers For more information see www jettcosmetics cz 13 5 The De...

Страница 26: ...or A 0001 15 DATE OF MANUFACTURE 15 2 2015 Compex spol s r o Palack ho t da 105 612 00 Brno Czech Republic tel 00420 602 766 759 www jettmedical cz Golden applicator with a precise tip B 0001 15 DATE...

Страница 27: ...27 C 0001 15 DATE OF MANUFACTURE 15 2 2015...

Отзывы: