Preparatory heating of skin
–
Caution:
The temperature of skin cannot exceed
41°C, which is done by the trained staff by controlling the skin with touch
thermometer.
Protection of skin from electric discharge
–
Caution:
In case of cosmetic treatments
(Intensity of discharge 1-5), a sufficient layer of indifferent gel must be applied on
skin. Application tip must be lead closely above the application gel. The optimal
distance during application is 1-2 mm. After thorough consideration, a doctor can use
a conductive serum instead of conductive indifferent gel.
Treatment close to the area of the eyes
–
operation with sharp tip.
Caution:
It is
necessary to carry the treatment in a safe distance from the eyes of patient. The trained
staff must make all possible precautions to protect eyes of his patient. Application
must be done in an absolute stillness and any other people must not be let into the
closeness of the treatment, as they could distract the operator. The trained staff takes
all responsibility for not using the sharp tip against the instructions recieved during the
training.
Proper treatment
–
Caution:
The trained staff is permitted to treat only the
recommended areas (see point 1.1
–
Use).
Grounding
–
Caution:
Before the treatment begins, the patient must be grounded by
a grounding disposable electrode which is stuck on the right shoulder blade.
Do not touch the tip
–
Caution:
Neither operator nor the patient are allowed to touch
the tip during the treatment. When the application tip gets away from the conductive
gel further than 5mm for more than 20 seconds, the supply of the tip is finished (the
yellow LED control FLASH goes off).
The voltage at the tip disappears in 1 second.
Do not touch the patient
–
Warning:
The trained
staff is not allowed to touch the
patient conductively. If there is a need to touch patient, it is necessary to use rubber
protective gloves.
Changing of application tip
–
Danger:
When the application tip is being used, Jett
Plasma Lift Device MUST BE unplugged from the power source. The device has 3
applicators
–
a spike, a flat applicator and a tip with a slim spike.
The trained staff can use the device only in the range defined by this manual ort
the qualification training.
Danger:
The trained person is allowed to use only the
equipment supplied by the manufacture. It is forbidden to open the device, because
there is a risk of electric shock.
After every use, the tip must be cleaned and disinfected.
Caution:
Protect the
device from hits and shocks.
Working with power adaptor.
Danger:
Do not pull the electric cable by force and
do not touch it with wet or moist hand.