SEM manual ETHZ v1.4
33
I.3) Tips on image acquisition time & magnification
•
Always use aperture 2
(or 1)
when acquiring images.
•
You can try to increase/decrease the current and adjust the counting time. A
medium
current
is recommended for most applications (spotsize 70-80) as it gives a
strong
signal
and a
decent sub-micron imaging resolution
. However:
o
A
lower current
(spotsize 50-70) will yield a
weaker signal
(
è
longer imaging
time required) and provide a
higher spatial resolution
(< 1 µm).
o
A
higher current
(spotsize >80) will
significantly increase the signal
but will
yield a
lower spatial resolution
(large beam size) and is
NOT recommended
for imaging
.
o
NOTE: If a super-high spatial resolution in SE mode is required (e.g., for imaging foraminifera,
small 3D-image of crystal), use a higher voltage (20 or 25 keV) and the beam aperture 1!
•
Prefer a medium to high magnification (> 100x) to avoid problems of barrel effect on
the edges or shadow effect from the BSE or CL detector.
•
For mosaic images, select a magnification sufficient to reveal all the feature in each
individual image (i.e., consider the true pixel size in micrometre). Check the
Appendix A4
for determining the pixel size of your individual image of the mosaic.
Do not choose a too high magnification that would require to acquire way too many
images! For instance, a magnification between ~70 and 150x is commonly used for
a mosaic image of a full thin section with large grains (> 50-100 µm), but a higher
magnification would be required to reveal micron-sized phases in smaller areas.
I.4) Exporting your images
Images are stored immediately when the acquisition is over. However, you will still need to
export individual images
to ultimately have an image with a scale bar (and with the point
you analysed in “Point & Shoot” shown on the image [see
Section J.3
]):
•
Select the image (or Point & Shoot) to export.
•
Click on menu “File > Export as FullRes TIF…”.
•
Click anywhere in the list of the acquired images. A new file ending with “…24bit.tif”
will appear. NOTE: This image is not readable in Thermo NSS; a grey image will
appear if you select it, that’s “normal”.
For mosaic images
, you will first need to MERGE all the files by selecting the menu “Batch
Processing > Create Montage”. Export then the resulting merged image as a full resolution
TIF image (menu “File > Export as FullRes TIF…” as described above).
Содержание JSM-6390 LA
Страница 2: ......
Страница 9: ...SEM manual ETHZ v1 4 5 Figure 5 Overview of the main buttons in the TOP section of the SEM program...
Страница 12: ...8 J M Allaz March 14 2021 Figure 9 Preparing the sample holder for thin section or for 1 round mount...
Страница 14: ...10 J M Allaz March 14 2021 Figure 10 Opening the sample chamber to remove or place a sample...
Страница 15: ...SEM manual ETHZ v1 4 11 Figure 11a Loading a new sample and taking an overview image of your sample SNS...
Страница 16: ...12 J M Allaz March 14 2021 Figure 11b Evacuating pumping the sample chamber after un loading a sample...
Страница 18: ...14 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Страница 23: ...SEM manual ETHZ v1 4 19 Figure 13 Complete procedure for beam alignment...
Страница 25: ...SEM manual ETHZ v1 4 21 Figure 15 Detail on the beam alignment procedure to obtain the best image quality...
Страница 30: ...26 J M Allaz March 14 2021 Figure 18 Creating a new NSS project or opening an existing one...
Страница 32: ...28 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Страница 34: ...30 J M Allaz March 14 2021 Figure 20 Electron Imaging mode acquire a single image in Thermo NSS...
Страница 36: ...32 J M Allaz March 14 2021 Figure 21 Electron Imaging mode acquire a mosaic image in Thermo NSS...
Страница 38: ...34 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Страница 41: ...SEM manual ETHZ v1 4 37 Figure 23 Loading standards into your project for quantitative EDS analysis...
Страница 43: ...SEM manual ETHZ v1 4 39 Figure 24 Spectrum Acquiring a single EDS spectrum over the currently scanned area...
Страница 45: ...SEM manual ETHZ v1 4 41 Figure 25 Point Shoot Acquiring multiple EDS analyses selected on an electron image...
Страница 48: ...44 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Страница 52: ...48 J M Allaz March 14 2021 Figure 27 Processing and extracting element maps...
Страница 53: ...SEM manual ETHZ v1 4 49 Figure 28 Calculating and extracting phase maps...
Страница 62: ...58 J M Allaz March 14 2021 A6 Thermo NSS toolbars from the NSS manual...