SEM manual ETHZ v1.4
9
C) Sample loading or unloading
The sample change procedure is summarised below. Refer also to
Figures 10 and 11a,b
.
C.1) Opening the sample chamber
1. In the JEOL SEM program, set the signal to SEI in the bottom section of the program.
2.
If the CL detector was in use, turn it OFF
(red button on the CL control box).
3. Turn ON the Deben infrared camera (if not already on).
4. Open the window “Stage” and move the stage to Z = 30 mm (absolute position).
5.
If the CL is inserted
(check on the IR camera!), you MUST remove the CL detector
(see
Figures 12 and 13
).
6. On the SEM column, close the valve VT-3 (tight, with the strength of two fingers).
7. Back to the JEOL SEM program, open the window “Sample” and click “Vent”.
WARNING: When the SEM is in
LaB
6
mode, check the vacuum gauge while venting
(see on the left side of the SEM) to ensure you don’t lose vacuum. This can sometime
happen due to the valve VT-3 leaking. Only the light “column” should turn orange, all other
lights should stay green.
If you lose vacuum, ask immediately for assistance!
This warning does NOT apply if the SEM is in
W-mode
(see
Appendix A2
).
8. When fully vented, open the sample chamber (see C.2 and
Figure 11
).
C.2) Removing/Loading a sample & closing the chamber
1. Open the sample chamber up to the back stop (use two hands!)
2. Remove the sample and/or load your sample (thin section or 1’’ round mount)
3. If you are loading a new sample, take a snapshot with SNS…
a. Place your sample on the holder.
b. Click the button “SNS” in the top section.
c. Click “Snap”. The stage will drive to the centre of the optical camera.
d. Wait a few seconds until a snapshot of your sample appears in the SEM
program. If you think the SNS image is not properly centred, refer to the
calibration section to calibrate the SNS in
Appendix A1
.
e.
IMPORTANT!
Lower again the stage to Z = 30 mm after using SNS!
4. Close the sample chamber while looking simultaneously at the IR camera to avoid a
disaster (e.g., sample holder touching a detector).
5. Open the window “Sample” and click “Evac” while pressing on the sample chamber
door. Wait a few seconds until the vacuum catches up.
6.
Wait until the signal “Ready” appears.
If this is the end of your session, you are done. Otherwise…
7. Open valve VT-3.
8. Remove the “Beam blank” and proceed with the gun activation (
Section E
) and the
beam alignment (
Section F
).
Содержание JSM-6390 LA
Страница 2: ......
Страница 9: ...SEM manual ETHZ v1 4 5 Figure 5 Overview of the main buttons in the TOP section of the SEM program...
Страница 12: ...8 J M Allaz March 14 2021 Figure 9 Preparing the sample holder for thin section or for 1 round mount...
Страница 14: ...10 J M Allaz March 14 2021 Figure 10 Opening the sample chamber to remove or place a sample...
Страница 15: ...SEM manual ETHZ v1 4 11 Figure 11a Loading a new sample and taking an overview image of your sample SNS...
Страница 16: ...12 J M Allaz March 14 2021 Figure 11b Evacuating pumping the sample chamber after un loading a sample...
Страница 18: ...14 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Страница 23: ...SEM manual ETHZ v1 4 19 Figure 13 Complete procedure for beam alignment...
Страница 25: ...SEM manual ETHZ v1 4 21 Figure 15 Detail on the beam alignment procedure to obtain the best image quality...
Страница 30: ...26 J M Allaz March 14 2021 Figure 18 Creating a new NSS project or opening an existing one...
Страница 32: ...28 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Страница 34: ...30 J M Allaz March 14 2021 Figure 20 Electron Imaging mode acquire a single image in Thermo NSS...
Страница 36: ...32 J M Allaz March 14 2021 Figure 21 Electron Imaging mode acquire a mosaic image in Thermo NSS...
Страница 38: ...34 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Страница 41: ...SEM manual ETHZ v1 4 37 Figure 23 Loading standards into your project for quantitative EDS analysis...
Страница 43: ...SEM manual ETHZ v1 4 39 Figure 24 Spectrum Acquiring a single EDS spectrum over the currently scanned area...
Страница 45: ...SEM manual ETHZ v1 4 41 Figure 25 Point Shoot Acquiring multiple EDS analyses selected on an electron image...
Страница 48: ...44 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Страница 52: ...48 J M Allaz March 14 2021 Figure 27 Processing and extracting element maps...
Страница 53: ...SEM manual ETHZ v1 4 49 Figure 28 Calculating and extracting phase maps...
Страница 62: ...58 J M Allaz March 14 2021 A6 Thermo NSS toolbars from the NSS manual...