26
= Key1
= Key 2
= Key 3
= Max. value or consumption = Min. value or supply
Energy measurement
In the UMG505, all in all 30 energy meters are at
your disposal. 24
meters can be cont-
rolled by tariff chan-
geover. When the
device is delivered,
12 energy meters are
displayed in the
measured value indi-
cation.
M VArh
kVArh
VArh
L1
L2
L3 ind
T
Press key 2 again.
The time of deletion
for real energy
is
indicated.
Year =01, Month =09
Day =06, Hour =08
Minute =15, Sec.=41
Time of deletion
For each energy meter, the running time is saved. If
real energy or reactive energy is deleted, all
corresponding tariffs are deleted as well. The time
of deletion is saved and running time is started
again.
As all real and reactive energy can be deleted
simultaneously, there is one time of deletion only for
all real and reactive energy meters.
The time of deletion can be called up directly within
the measured value indications as additional
information for the energy meters, provided this
energy meter is configured for measured value
indication (see manufacturer settings).
Example: Call up deletion time for real energy
The time of deletion can be called up in the
measured value indication of real energy. To reach
the first measured value indication from each
programming part, press key 1 for about 2 seconds.
M Wh
kWh
Wh
L1
L2
L3
T
SELECT
Y.M
D.H
M.S
Chose select mode
using key 1.
The symbol
SELECT
flashes.
Confirm with key
2.
The symbol
SELECT
is on.
SELECT
M Wh
kWh
Wh
L1
L2
L3
T
When the device is delivered, only the grey meters
can be called up within the measured value indicati-
on.
Energy meter
Changeable
Real energy
Without rev. run. stop T50 T51 T52 T53 T54
Consumption
T00 T01 T02 T03 T04
Supply
T30 T31 T32 T33 T34
Reactive energy
Without rev. run. stop T40 T41 T42 T43 T44
inductive
T10 T11 T12 T13 T14
capacitive
T20 T21 T22 T23 T24
Diagr. Overview energy meters
Scroll to measured
value indication of
real power T00 by
pressing key 3.
Press
key 1 for about 2 seconds and you return to
the first measured value window of the measured
value indication from each program part.