GO-5000M-PGE-UV User Manual (Ver. 1.2)
Main Functions
- 52 -
Multi-Frame
In this mode, the AcquisitionStart command captures the number of frames which are specified by
AcquisitionFrameCount.
◆
Normal multi-frame operation
1. AcquisitionStart command is input
2. AcquisitionTriggerWait becomes effective
3. AcquisitionActive becomes “TRUE” (accepts capture)
4. Output N frames as specified by AcquisitionFrameCount
5. AcquisitionActive becomes “FALSE”. Then the output stops. (See the following diagram)
Note:
This figure is if the trigger is set to ON. When the trigger is OFF, FrameActive becomes True
at the same timing of AcquisitionActive.
◆
Forcing acquisition to stop
While AcquisitionActive is “TRUE”, if AcquisitionStop or AcquisitionAbort is initiated, AcquisitionActive
becomes “FALSE” (stop capturing). Once the operation is set to “FALSE”, the internal FrameCount is
reset.
However, if AcquisitionStop command is initiated during image output period, AcquisitionActive
becomes “FALSE” (stop capturing) after image output is completed.
Once, AcquisitionActive becomes “FALSE”, the internal count is reset.
Acquisition Frame Count: Can be set in the range of 1 to 255
Содержание GO-5000M-PGE-UV
Страница 46: ...GO 5000M PGE UV User Manual Ver 1 2 Main Functions 46 Output Timing Vertical Vertical Timing ...
Страница 55: ...GO 5000M PGE UV User Manual Ver 1 2 Main Functions 55 How to Determine the Maximum Frame Rate ...
Страница 60: ...GO 5000M PGE UV User Manual Ver 1 2 Main Functions 60 ALC Channel Area ...
Страница 74: ...GO 5000M PGE UV User Manual Ver 1 2 Main Functions 74 Multi ROI Setting Example ...
Страница 86: ...GO 5000M PGE UV User Manual Ver 1 2 Miscellaneous 86 Spectral Response ...