background image

Therefore, H

v

 is 

 
 

(4). 

 
 
 
Magnetic flux density B

p

 inside a single sheet is 

 

(5). 

 
 
A conventional MC method-based SST is equivalent to measuring the iron loss  P per unit volume of a single 
sheet  from  this  H

v

  and  B

p

(Whenever  “iron  loss”  is  indicated  hereafter,  it  mean  iron  loss  per  unit  volume.) 

Because iron loss P is determined by dividing the area of the B-H curve by the cycle of the excitation signal, 
 

(6), 

 
when formula (2) is substituted into this, it becomes 
 
 
 

(7). 

 
 
 
The first term of formula (7) is the actual iron loss P

p,

 which only includes the magnetic field strength H

p

 and the 

magnetic flux density B

p

 in a single sheet. The second term is related to the iron loss H

ya

 in the measuring yoke, 

which  includes  the  magnetic  field  strength  in  the  yoke.  A  conventional  MC  method-based  SST  ignores  the 
second  term  of  formula  (7)  by  assuming  that  the  magnetic  field  strength  in  a  measuring  yoke  H

ya

  is  small 

compared to the magnetic field strength in a single sheet H

p

 (H

v

 

≈ H

p

) in formula (2). 

 
With  this  instrument,  magnetic  field  strength  in  the  measuring  yoke  H

ya

,  which  was  unmeasurable  with 

conventional  MC  method-based  SSTs,  can  be  calculated  from  this  1st  measurement  and  the  2nd  and  3rd, 
explained later. Using the value of H

ya

, find the magnetic field strength H

p

 of the single sheet and then calculate 

the actual iron loss P

p

 of the single sheet with only the first term of formula (7).   

 
In the 1st measurement, H

v

 and B

p

 are calculated by formulas (4) and (5). 

 
The 2nd measurement is explained next. 
 
Reconnect Relay 2 to iii-side. Do not change Relays 1 and 3. From the signal generator, send an excitation sine 
wave signal of the same frequency f (cycle T) used in the 1st measurement so that the excitation current i

1a2

which flows through the excitation coil on the measuring yoke, is the same as the excitation current i

1a

 from the 

1st measurement. At this point, an induced voltage V

3a

 is generated at both ends of the B coil on the measuring 

yoke. 
 
Magnetic flux density B

ya

 in the measuring yoke is 

 

(8). 

 

 
 

 

H

v

=

n

1

 i

1a

L

p

 

    =

n

1

 V

1a

L

p

 R

s

 

 
 

B

p

=

1

n

2

 A

p

∫ V

2

dt 

 
 

P =

1
T

∫ H

v

dB

p

dt

dt

T

0

 

 
 

P =

1
T

∫ {H

p

+ (

L

ya

L

p

) H

ya

}

dB

p

dt

dt

T

0

 

    =

1
T

∫ H

p

dB

p

dt

dt

T

0

+

1
T

∫ (

L

ya

L

p

) H

ya

dB

p

dt

dt

T

0

 

 
 

B

ya

=

1

n

3

 A

y

∫ V

3a

dt 

 
 

Содержание SY-956

Страница 1: ...Instruction Manual Mini Single Sheet Tester SY 956...

Страница 2: ...2014_2015 IWATSU TEST INSTRUMENTS CORPORATION All rights reserved...

Страница 3: ...operation or failure to observe the warning may result in death or serious injury Cautions Incorrect operation or failure to observe the caution may result in injury or damage to instrument Notes Par...

Страница 4: ...is product if wet otherwise an electric shock or fire could occur If water gets on or inside this product turn off the main power switch on the front side and remove the power plug from the receptacle...

Страница 5: ...d since this product automatically adapts to the power supply voltage Strictly observe items below when handling the power cord If not an electric shock or fire may occur If the power cord is damaged...

Страница 6: ...y cause an electric shock fire or failure If water chemical metal object gets in turn off the main power switch on the front side and remove the power plug from the receptacle Then contact Iwatsu offi...

Страница 7: ...fire may occur In addition if a 2 core power cord is used a fire may occur Unless specified when purchasing this product the power cord suitable for 100V system center voltage 100V to 120V is attached...

Страница 8: ...hock may occur Do not use any damaged cable or power cord Otherwise an electric shock or fire may occur If this product will not be used for a long time remove the power plug from the receptacle for s...

Страница 9: ...or an item damaged during transportation immediately contact Iwatsu office or our sales distributors Components Mini Single Sheet Tester SY 956 main unit 1 Accessories Mini Single Sheet Tester cable...

Страница 10: ...ervice center Any failure which occurs within the term of warranty and for which Iwatsu is responsible should be repaired without any cost When shipping a product to be repaired clearly write the prod...

Страница 11: ...ing Mini Single Sheet Tester Cable SY 957 16 4 4 Powering on 18 4 5 Powering off 18 4 6 Attaching and detaching a B Coil 19 4 7 Measurable samples 20 4 8 Setting a sample 21 4 9 Confirmation of functi...

Страница 12: ...iction is used for the measuring yoke 3 Highly accurate measurement of single sheet core loss has been achieved by providing the Yoke Compensation 1 function which cancels the measurement yoke s magne...

Страница 13: ...e of temperature and humidity If exposed to direct sunlight or used outside an operating range e g high humidity it may fail Ranges of temperature and humidity for usage are as follows Indoor use only...

Страница 14: ...measurement using the same B magnetic flux density coil Although 2 types of B coils are provided with this instrument it is the best practice for users to wind a B coil by themselves with considerati...

Страница 15: ...ing 6 Conduct a comparative measurement under the same temperature It is a basic practice to conduct a comparative measurement under the same temperature as magnetic properties of some materials such...

Страница 16: ...in the case of SSTs As such Pcv measured without any sample is the detection limit of core loss Pcv of the particular device with the same frequency under the same excitation conditions That means th...

Страница 17: ...yoke Core area Ay Relative permeability y equipped with an excitation coil Number of turns n1 and a B coil Number of turns n3 a signal generator that sends excitation current to each excitation coil...

Страница 18: ...h this instrument magnetic field strength in the measuring yoke Hya which was unmeasurable with conventional MC method based SSTs can be calculated from this 1st measurement and the 2nd and 3rd explai...

Страница 19: ...e double yoke is 9 When the magnetic path length of the double yoke is Lyb the magnetic field strength Hyb in the double yoke can be calculated from Ampere s Law 10 Hyb is the magnetic field strength...

Страница 20: ...2nd measurement this instrument controls the amplitude of the signal generator precisely and adjusts the excitation current i1b sent through the double yoke However the generator cannot be controlled...

Страница 21: ...hed Operation is invalid if the button is pressed while the LED is blinking 5 SET button Pressing this button when the LED is off will cause the slider to move down while the LED blinks and press a sa...

Страница 22: ...ock is made of poly phenylene sulfide PPS resin PPS resin may deform at 160 C or higher 3 Slider Move the press block up and down 4 Guide arms Support long samples from the side Guide arms are made of...

Страница 23: ...ect the output from the power amplifier with the BNC cable 3 Connection terminals for Mini Single Sheet Tester Cable SY 957 Connect to the B H Analyzer with Mini Single Sheet Tester Cable SY 957 1 Pow...

Страница 24: ...the B H Analyzer See sec 4 10 Select the single sheet testing mode Select Sheet in the Mode menu of the B H Analyzer See sec 4 11 Set up a sample on SY 956 See sec 4 8 Measurement See sec 4 15 Warm u...

Страница 25: ...eeds to be connected to the B H Analyzer with which it has been calibrated together Proper measurement cannot be made by connecting to any other B H Analyzers Fig 4 2 Connecting SY 956 B H Analyzer OS...

Страница 26: ...ews so that they do not get lost 4 Install the measuring block for Mini Single Sheet Tester Cable SY 957 First align the terminal block guide with the measuring POD guide See Fig 4 5 4 6 Then slowly p...

Страница 27: ...et Tester Cable SY 957 on the back of the product Fit the 4 BNCs and a multi pin connector together See Fig 4 8 7 Secure the connector by turning the lever to the right See Fig 4 9 Fig 4 8 Connecting...

Страница 28: ...it start up Refer to the B H Analyzer s instruction manual for the operation 5 Warm it up for 30 minutes or more 4 5 Powering off The following is the method to power off this instrument Follow the po...

Страница 29: ...S2 terminals Place the excess wire in the concavity of the press block to prevent interference with the press block and the measuring yoke when the slider moves up and down Do not modify the provided...

Страница 30: ...measurable conditions of samples Conditions of samples Expected failures Wet samples Moisture will be attached to the yoke causing it to rust Dirty samples Dirt will be attached to the yoke causing it...

Страница 31: ...ue of the Slider Position No P with the conversion formula given in the table When the thickness of a sample is T 0 7 mm and the force with which to press the sample is F 5 7 N Substitute T and F into...

Страница 32: ...to display the Mode menu 2 Move the editing cursor to Sheet press and confirm selection of the single sheet mode See Fig 4 16 3 The single sheet testing mode screen is displayed See Fig 4 18 4 11 Conf...

Страница 33: ...Shows whether the currently displayed graph is Current current measurement or Reference stored measurement 5 Cursor Values Shows the cursor value when the cursor is displayed Unit is same as the grap...

Страница 34: ...ules Description Sample Name Up to 20 characters Le Effective magnetic path length System 2 Device specific eigenvalue Device specific size inside the poles of the measuring yoke Unit is fixed to mm U...

Страница 35: ...Hz Set the frequency of an excitation signal Fixed Parameter Target parameter Bm Max magnetic flux density 0 001 to 99999 Set a target parameter value for measurement Hm Max magnetic field I1m Max exc...

Страница 36: ...orner of the B H graph screen the top left corner on the time axis graph to show that the third measurement is taking place See Fig 4 20 4 When the 3rd measurement has been completed successfully the...

Страница 37: ...27 Fig 4 19 1st Measurement Screen Fig 4 20 3rd Measurement Screen...

Страница 38: ...ase refer to the B H Analyzer s instruction manual for the operation 4 20 USB memory output of screen hardcopy Similar to the normally installed testing modes output of a measuring screen hardcopy to...

Страница 39: ...per or the like This would cause an increase in core loss of the yoke leading to incorrect measurement 2 Lightly wipe the measuring yoke except its pole faces with a soft cloth moistened with a small...

Страница 40: ...e Comp Enabled 4 0 15 deg Typical f 10 kHz 200 mA and 200 mV range or more at the amplitude of 80 or more of the range in use Core loss Yoke Comp Disabled 3 5 6 Typical Estimated value at phase angle...

Страница 41: ...31 6 2 Outside appearance diagram Unit mm...

Страница 42: ...32 6 3 Outside appearance of the fixture Unit mm...

Страница 43: ...Address 7 41 Kugayama 1 chome Suginami ku Tokyo 168 8511 Japan Phone 81 3 5370 5483 Facsimile 81 3 5370 5492 Homepage http www iti iwatsu co jp...

Страница 44: ...SY 956...

Отзывы: