10
In conclusion, by substituting H
p
of formula (12) into the first term of formula (7),
(13),
the actual iron loss of only the single sheet P
p
can be calculated.
Yoke Comp. comes into effect when Yoke Comp. is Enabled, as explained in
4.15. When Yoke Comp. is
Disabled, testing is completed after the 1st measurement.
Still,
Yoke Comp. does not necessarily work at all times
. In order to make the magnetic flux density B
ya
of
the double-yoke in the 3rd measurement equal to the magnetic flux density B
ya
of the measuring yoke in the
2nd measurement, this instrument controls the amplitude of the signal generator precisely and adjusts the
excitation current i
1b
sent through the double-yoke. However, the generator cannot be controlled when this B
ya
is extremely small. When a warning message is displayed and the 3rd measurement does not proceed after the
2nd measurement, it is usually the case that this B
ya
is extremely small (i.e., measurement conditions are such
that the core loss related to the measuring yoke is extremely small).
Fig. 2-4 Summary of the measurement principles
P
p
=
1
T
∫ H
p
dB
p
dt
dt
T
0
Signal
generator
Shunt resistor
Relay 1
B Coil
Single sheet
Excitation
coil
Measuring yoke
Double-yoke
Relay 2
Relay 3
B Coil
B Coil
Excitation
coil
Содержание SY-956
Страница 1: ...Instruction Manual Mini Single Sheet Tester SY 956...
Страница 2: ...2014_2015 IWATSU TEST INSTRUMENTS CORPORATION All rights reserved...
Страница 37: ...27 Fig 4 19 1st Measurement Screen Fig 4 20 3rd Measurement Screen...
Страница 41: ...31 6 2 Outside appearance diagram Unit mm...
Страница 42: ...32 6 3 Outside appearance of the fixture Unit mm...
Страница 44: ...SY 956...