2.2   Health and Safety Instructions

This furnace has been designed according to EN 61010-1 and has 
been shipped from the manufacturer in excellent condition as far as 
safety regulations are concerned. To maintain this condition and to 
assure risk-free operation, the user must observe the notes and 
warnings contained in these Operating Instructions.

–  The user must especially become familiar with the warnings and 

the operating conditions to prevent injury to personnel or damage 
to materials. The manufacturer is not responsible for damage 
resulting from misuse or failure to observe the Operating Instruc-
tions. Warranty claims cannot be accepted in such cases.

–  Before switching on the furnace, make sure that the voltage  

indicated on the rating plate complies with your local power  
supply.

–  The power socket must be equipped with a residual current circuit 

breaker. 

–  The furnace must be plugged into a socket with protected  

contacts.

–  Place furnace on a fire-proof table (observe local regulations, e.g. 

distance to combustible substances or objects, etc.)

–  Always keep the air vents at the rear and the side of the furnace 

free from obstruction.

–  Do not touch any parts that become hot during the operation of 

the furnace. There is a burn hazard!

–  Clean furnace only with a dry or slightly moist cloth. Do not use 

any solvents! Disconnect power before cleaning.

–  The furnace must be cool before it is packed for transportation 

purposes.

–  Use original packaging for transportation purposes.

–  Before calibration, maintenance, repair, or exchange of parts, the 

power must be disconnected if the furnace is to be opened.

–  If calibration, maintenance, or repair has to be carried out with 

the power connected and the furnace open, only qualified 
 personnel, who are familiar with the risks and dangers, may 
 perform these procedures.

–  After maintenance, the required safety tests (high voltage  

resistance, protective conductor, etc.) have to be carried out.

–  Ensure that only fuses of the indicated type and rated current are 

used.

–  If it is assumed that safe operation is no longer possible, the  

power must be disconnected to avoid accidental operation.

  Safe operation is no longer possible if
  –  the furnace is visibly damaged
  –   the furnace does not work
  –   the furnace has been stored under  

  unfavourable conditions over an extended period of time

–  Use only original spare parts.

–  The temperature range for faultless operation is +5 °C to +40 °C 

(+41 °F to +104 °F).

–  If the furnace has been stored at very low temperatures or high 

atmospheric humidity the head has to be opened and the unit 
dried or left to adjust to room temperature for approx.  
1 hour (do not connect the power yet).

–  The furnace has been tested for use at altitudes of up to 2000 m 

(6562 ft) above sea level.

–  The furnace may only be used indoors. 

  Any disruption of the protective conductor either inside 

or outside the furnace or any loosening of the protective 
conductor connection may lead to danger for the user in 
case of malfunction. Deliberate interruptions are not  
tolerated. 

  Materials developing harmful gases must not be fired.

Warnings regarding the dismounting of the heating muffle

This product contains ceramic fibres and may release fibre 
dust. Fibre dust has proved to be carcinogenic in animal 
experiments. The corresponding EU Safety Data Sheet must 

be observed.

 
The heat insulation of the firing chamber in the Programat CS  
consists of ceramic fibres. After prolonged use of ceramic fibres at 
temperatures of over 900 °C (1652 °F), silicogenic substances  
(Cristobalite) may be produced. In certain cases, e.g. upon changing 
of the heating muffle, the possible resulting dust exposure may 
cause irritation of the skin, eyes, and respiratory organs. Therefore, 
proceed as follows when changing the heating muffle:

 
–  Make sure the corresponding staff wears long-sleeved clothing, as 

well as headgear, goggles, and gloves.

–  Place suction equipment at the source of the dust or, if not  

possible, provide the staff with FFP3 facemasks or similar items.

–  Once the procedure has been completed, any dust possibly  

adhering to exposed skin must first be rinsed off with cold water. 
Only after that should soap and warm water be used.

–  The corresponding work clothes should be washed separately.

Warning 
The insulation on this product contains refractory ceramic fibres (RCF) 
which pose a possible cancer hazard, if agitated and inhaled. May be 
irritating to the skin, eyes or respiratory tract if insulation is cracked 
or corrupted. 

California Proposition 65  
Warning: ”This product contains Refractory  
Ceramic Fibres, a substance known to the State of California to 
cause cancer.” 

Disposal:

The furnaces must not be disposed of in the normal 
domestic waste. Please correctly dispose of old 
 furnaces according to the corresponding EU council 
directive. Information on the correct disposal may 
also be found on your local Ivoclar Vivadent home-
page.

Содержание Programat CS

Страница 1: ...Operating Instructions Programat CS Valid as of Software Version 3 0...

Страница 2: ...2...

Страница 3: ...key functions 5 3 Basic meaning of the display information 5 4 Program structure 5 5 Adjustable parameters and possible value ranges 5 6 Settings test programs and information 5 7 Explanation of the...

Страница 4: ...20 Thermocouple cable 21 Vacuum hose connection 23 Rubber feet 24 Firing plate holder 25 Furnace head housing 26 Thermocouple plug 27 Plug fuse 28 Heater plug 29 Heater plug socket 30 Thermocouple pl...

Страница 5: ...5 40 25 34 6 10 35 23 47 33 1 7 5 2 3 24 4 8 9...

Страница 6: ...6 20 26 30 29 48 49 27 28 11 32 43 37 39 36 11 38 19 13 12 12 13 15 19 46 21 17 18 16 25...

Страница 7: ...right 80 Cursor key left 81 Information key 82 Program 1 83 Program 2 84 Program 3 85 Program 4 86 Next program 87 Open furnace head 88 Close furnace head 8 4 44 3 100 Programat firing tray 101 Metal...

Страница 8: ...ave the following meanings Risks and dangers Important information Contraindication Burn hazard Risk of crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Instructions...

Страница 9: ...t be observed The instructions regulations and notes in the material s Instructions for Use must be observed The furnace must be operated under the indicated environmental and operating conditions Cha...

Страница 10: ...hazard Never place objects in the firing chamber by hand since there is a burn hazard Always use the tongs accessories supplied for this pur pose Never touch the hot surface of the furnace head as th...

Страница 11: ...original spare parts The temperature range for faultless operation is 5 C to 40 C 41 F to 104 F If the furnace has been stored at very low temperatures or high atmospheric humidity the head has to be...

Страница 12: ...ctronic controls and a software Moreover the set and actual temperatures are continuously compared 3 4 Accessories not part of the delivery form Automatic Temperature Checking Set 2 ATK 2 Programat Ac...

Страница 13: ...Simply use the two corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed with the regular household refuse...

Страница 14: ...5 on the firing plate holder 24 Secure the cooling tray 34 with the two screws 35 including the silicone washer 47 Place the cooling tray 34 on the frame plate 7 Make sure that the cooling tray 34 is...

Страница 15: ...nace head mounting Ensure that the furnace head mounting mark 41 is aligned with the furnace base mounting mark 42 Make sure that the firing plate 5 is not damaged by mounting the furnace head Step 4...

Страница 16: ...itional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord 16 with the power socket 17 of the furn...

Страница 17: ...with the wall socket 2 Put the On Off switch 11 at the rear of the furnace on position I The furnace will now automatically conduct a self test The performance of all furnace components is automatica...

Страница 18: ...rogram Starting a program is only possible with the furnace head open Start LED 74 Illuminated if a program has been started The LED is blinking while a program is paused STOP key 75 Pressing this key...

Страница 19: ...ration and information 5 5 Adjustable parameters and possible value ranges Symbol Parameter Value range Value range P Program number P 1 20 B Stand by temperature 100 700 C 212 1292 F S Closing time m...

Страница 20: ...by means of the or keys The tune can be set by means of the or keys Enables the pro gramming of the parameters of the currently selected program Settings Renaming Time Date General write protection V...

Страница 21: ...e Ivoclar Vivadent optimized temperature control function With this setting all values and para meters can be reset to the factory settings Attention All individual pro grams which have been created a...

Страница 22: ...fic program two stage program is used The bold line indicates the values for the second stage Furnace head Is shown in the recommendation open area and indicates the most likely action to follow Furna...

Страница 23: ...m Step 1 Select a free program Step 2 Select the parameter list via the setting Programming 7 20 and change the program parameters using the keys After that change back to the Stand by screen by press...

Страница 24: ...e program is used the parameters are displayed for the first and the second stage in a separate screen Switching between the two screens is carried out using the symbol Change program stage indication...

Страница 25: ...lug in connections for correct fit Var external connections weekly Check if the furnace head opens smoothly and without excessive noise Opening mechanism monthly Check if the thermocouple is straight...

Страница 26: ...mend keeping the furnace head closed especially if the temperature drops below 150 C 302 F 7 5 Dehumidification program The condensation of water in the insulation of the firing chamber and the vacuum...

Страница 27: ...o hot to start a firing program start of a firing program 17 System Furnace head cannot 27 The furnace head cannot be moved to the final position It might be blocked by an external mechanical be initi...

Страница 28: ...ales Service if one of the following error messages being displayed 25 29 32 43 44 45 46 47 48 54 56 103 107 700 701 703 704 705 706 707 800 1010 1011 1012 1013 1014 1015 1016 1024 1025 1026 1028 1202...

Страница 29: ...n and off again Has the vacuum already been released Is the program still running Wait until the pro gram is complete Switch furnace off and on again Vacuum pump does not start working Is the vacuum p...

Страница 30: ...120 V Diameter 6 3 x 32 mm 200 240 V Diameter 5 x 20 mm Dimensions of the closed furnace Depth 430 mm width 305 mm 410 mm with Cooling tray Height 320 mm Usable size of the firing chamber Diameter 80...

Страница 31: ...uld correspond to the Software version which is used on your furnace 10 2 Program structure Program Description 1 Crystallization glaze program for IPS e max CAD HT LT 2 Corrective firing for IPS e ma...

Страница 32: ...ga St Rosedale PO Box 303011 North Harbour Auckland 0751 New Zealand Tel 64 9 914 99 99 Fax 64 9 914 99 90 www ivoclarvivadent co nz Ivoclar Vivadent Polska Sp z o o Al Jana Pawla II 78 00 175 Warszaw...

Отзывы: