9. Product Specifications
9.1 Delivery form
– Programat CS
– Power cord
– Vacuum hose
– Cooling Tray “S”
– Programat Firing tongs
– Programat Firing Tray Kit
– Automatic Temperature Calibration Set ATK2 (Test Set)
– USB data cable
– PrograBase 2 Software
– Service passport equipment
– Operating Instructions
Recommended accessories
– Programat Accessories Set
– Automatic Temperature Checking Set 2 (ATK 2)
– Vacuum Pump VP3 easy
9.2 Technical data
Power supply
100 V / 50–60 Hz
110–120 V / 50–60 Hz
200–240 V / 50–60 Hz
Overvoltage category II
Contamination level 2
Tolerated voltage fluctuations
+/- 10%
Max. power consumption
11 A at 100 V
12 A at 110–120 V
8.5 A at 200–240 V
Acceptable data for vacuum pump of other manufacturers
Max. output:
250 W / max. leakage current 0.75 mA
Final vacuum:
< 50 mbar
Use only tested pumps
Electrical fuses:
100 V / 110–120 V:
250 V / T 15 A (heating circuit)
250 V / T 5 A (vacuum pump)
200–240 V:
250 V / T 8 A ( heating circuit)
250 V / T 3.15 A ( vacuum pump)
Dimensions of electrical fuses
110–120 V:
Diameter 6.3 x 32 mm
200–240 V:
Diameter 5 x 20 mm
Dimensions of the closed furnace
Depth: 430 mm / width: 305 mm / 410 mm (with Cooling tray)
Height: 320 mm
Usable size of the firing chamber
Diameter 80 mm
Height 48 mm
Max. firing temperature
1200 °C (2192 °F)
Weight
Furnace base: 12.0 kg
Furnace head: 4.5 kg
Safety information
The CS complies with the following guidelines:
– IEC 1010-1/EN 61010, Part 1
– UL and cUL standards
Radio protection / electromagnetic compatibility
EMC tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:
+5 °C to +40 °C (+41 °F to +104 °F)
Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31 °C
(87.8 ° F) gradually decreasing to 50 % relative humidity at 40 °C
(104 °F); condensation excluded.
Acceptable ambient pressure:
The furnace is tested for use at altitudes of up to 2000 m above sea
level.
9.4 Acceptable transportation and storage conditions
Acceptable temperature range
-20 to +65 °C (-4 °F to +149 °F)
Acceptable humidity range
Max. 80 % relative humidity
Acceptable ambient pressure
500 mbar to 1060 mbar
Use only original packaging of the Programat CS together with the
respective foam material for shipping purposes.