8.3   Technical malfunctions

These malfunctions may occur without an error message being displayed.
* If there are any questions, please contact the Ivoclar Vivadent After Sales Service.

8.4 Repair

  Repairs may only be carried out by a certified Ivoclar  

Vivadent Service Center. Please refer to the addresses on 
the last page of these Operating Instructions.

If repairs during the warranty period are not carried out by a certified 
Ivoclar Vivadent Service Center, the warranty will expire immediately. 
Please also refer to the corresponding warranty regulations.

Description

Double-check

Action

 

Vacuum is not released or only very slowly

Is the vacuum released within approximately  
30 seconds?

Wait until the vacuum is released, remove object.  
Switch the furnace on and off again * 

Indication on display incomplete

Activate the display test program 

*

Writing in the display is very hard to read

Is the contrast properly set?

Adjust contrast

Display not illuminated

Is the furnace properly connected according to the 
Operating Instructions and switched on?

Correctly connect the furnace and switch it on.

Buzzer does not sound

Is the buzzer switched off (Tune 0)?

Select tune 1–5

Furnace head does not open

Was the furnace head moved manually?

Open the furnace head only by using the  
corresponding keys. Switch the furnace on and off 
again

Has the vacuum already been released?

Is the program still running? Wait until the pro-
gram is complete. Switch furnace off and on again 
*

Vacuum pump does not start working

Is the vacuum pump fuse defective?

Check fuse and replace if necessary.

Was the maximum power consumption exceeded?

Use only the vacuum pump recommended by 
 Ivoclar  Vivadent.

Is the vacuum pump plug correctly connected?

Correctly connect the vacuum pump to the furnace 
base. 

Final vacuum is not reached

Is the vacuum hose OK?

Check vacuum hose and hose connection.

Is the pump output OK?

Start the vacuum test program.

Humidity/condensation in the vacuum hose?

Start dehumidification program.

Incorrect or illogical temperature indication

Is the thermocouple bent or fractured?

*

Is the thermocouple correctly connected?

Correctly connect thermocouple.

Is the thermocouple plug defective?

*

Hairline cracks in the heating muffle

Are the cracks very small and insignificant (hairline 
cracks)?

Small cracks in the muffle are normal and do not 
negatively influence the function of the furnace.

Are the cracks large or have parts of the heating 
muffle broken off?

*

Cracks in the insulation

Are the cracks very small and insignificant (hairline 
cracks)?

Small cracks in the insulation do not negatively 
influence the furnace.

Are the cracks large or have parts of the insulation 
broken off?

*

Cracks in the quartz glass / heating element

Are there cracks in the quartz glass or is the quartz 
glass sheathing the heating wire broken?

Switch off the furnace *

Содержание Programat CS

Страница 1: ...Operating Instructions Programat CS Valid as of Software Version 3 0...

Страница 2: ...2...

Страница 3: ...key functions 5 3 Basic meaning of the display information 5 4 Program structure 5 5 Adjustable parameters and possible value ranges 5 6 Settings test programs and information 5 7 Explanation of the...

Страница 4: ...20 Thermocouple cable 21 Vacuum hose connection 23 Rubber feet 24 Firing plate holder 25 Furnace head housing 26 Thermocouple plug 27 Plug fuse 28 Heater plug 29 Heater plug socket 30 Thermocouple pl...

Страница 5: ...5 40 25 34 6 10 35 23 47 33 1 7 5 2 3 24 4 8 9...

Страница 6: ...6 20 26 30 29 48 49 27 28 11 32 43 37 39 36 11 38 19 13 12 12 13 15 19 46 21 17 18 16 25...

Страница 7: ...right 80 Cursor key left 81 Information key 82 Program 1 83 Program 2 84 Program 3 85 Program 4 86 Next program 87 Open furnace head 88 Close furnace head 8 4 44 3 100 Programat firing tray 101 Metal...

Страница 8: ...ave the following meanings Risks and dangers Important information Contraindication Burn hazard Risk of crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Instructions...

Страница 9: ...t be observed The instructions regulations and notes in the material s Instructions for Use must be observed The furnace must be operated under the indicated environmental and operating conditions Cha...

Страница 10: ...hazard Never place objects in the firing chamber by hand since there is a burn hazard Always use the tongs accessories supplied for this pur pose Never touch the hot surface of the furnace head as th...

Страница 11: ...original spare parts The temperature range for faultless operation is 5 C to 40 C 41 F to 104 F If the furnace has been stored at very low temperatures or high atmospheric humidity the head has to be...

Страница 12: ...ctronic controls and a software Moreover the set and actual temperatures are continuously compared 3 4 Accessories not part of the delivery form Automatic Temperature Checking Set 2 ATK 2 Programat Ac...

Страница 13: ...Simply use the two corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed with the regular household refuse...

Страница 14: ...5 on the firing plate holder 24 Secure the cooling tray 34 with the two screws 35 including the silicone washer 47 Place the cooling tray 34 on the frame plate 7 Make sure that the cooling tray 34 is...

Страница 15: ...nace head mounting Ensure that the furnace head mounting mark 41 is aligned with the furnace base mounting mark 42 Make sure that the firing plate 5 is not damaged by mounting the furnace head Step 4...

Страница 16: ...itional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord 16 with the power socket 17 of the furn...

Страница 17: ...with the wall socket 2 Put the On Off switch 11 at the rear of the furnace on position I The furnace will now automatically conduct a self test The performance of all furnace components is automatica...

Страница 18: ...rogram Starting a program is only possible with the furnace head open Start LED 74 Illuminated if a program has been started The LED is blinking while a program is paused STOP key 75 Pressing this key...

Страница 19: ...ration and information 5 5 Adjustable parameters and possible value ranges Symbol Parameter Value range Value range P Program number P 1 20 B Stand by temperature 100 700 C 212 1292 F S Closing time m...

Страница 20: ...by means of the or keys The tune can be set by means of the or keys Enables the pro gramming of the parameters of the currently selected program Settings Renaming Time Date General write protection V...

Страница 21: ...e Ivoclar Vivadent optimized temperature control function With this setting all values and para meters can be reset to the factory settings Attention All individual pro grams which have been created a...

Страница 22: ...fic program two stage program is used The bold line indicates the values for the second stage Furnace head Is shown in the recommendation open area and indicates the most likely action to follow Furna...

Страница 23: ...m Step 1 Select a free program Step 2 Select the parameter list via the setting Programming 7 20 and change the program parameters using the keys After that change back to the Stand by screen by press...

Страница 24: ...e program is used the parameters are displayed for the first and the second stage in a separate screen Switching between the two screens is carried out using the symbol Change program stage indication...

Страница 25: ...lug in connections for correct fit Var external connections weekly Check if the furnace head opens smoothly and without excessive noise Opening mechanism monthly Check if the thermocouple is straight...

Страница 26: ...mend keeping the furnace head closed especially if the temperature drops below 150 C 302 F 7 5 Dehumidification program The condensation of water in the insulation of the firing chamber and the vacuum...

Страница 27: ...o hot to start a firing program start of a firing program 17 System Furnace head cannot 27 The furnace head cannot be moved to the final position It might be blocked by an external mechanical be initi...

Страница 28: ...ales Service if one of the following error messages being displayed 25 29 32 43 44 45 46 47 48 54 56 103 107 700 701 703 704 705 706 707 800 1010 1011 1012 1013 1014 1015 1016 1024 1025 1026 1028 1202...

Страница 29: ...n and off again Has the vacuum already been released Is the program still running Wait until the pro gram is complete Switch furnace off and on again Vacuum pump does not start working Is the vacuum p...

Страница 30: ...120 V Diameter 6 3 x 32 mm 200 240 V Diameter 5 x 20 mm Dimensions of the closed furnace Depth 430 mm width 305 mm 410 mm with Cooling tray Height 320 mm Usable size of the firing chamber Diameter 80...

Страница 31: ...uld correspond to the Software version which is used on your furnace 10 2 Program structure Program Description 1 Crystallization glaze program for IPS e max CAD HT LT 2 Corrective firing for IPS e ma...

Страница 32: ...ga St Rosedale PO Box 303011 North Harbour Auckland 0751 New Zealand Tel 64 9 914 99 99 Fax 64 9 914 99 90 www ivoclarvivadent co nz Ivoclar Vivadent Polska Sp z o o Al Jana Pawla II 78 00 175 Warszaw...

Отзывы: