Harmonic Measurement Commands
Copyright © Itech Electronic Co., Ltd.
39
FETCh[:SCALar]:HARMonics:POWer[:ACTive]:FUNDa
mental?
This command is used to read the latest pre-processing power fundamental
harmonic.It will not affect the device setting values and will not trigger a new
measurement operation.It only requires to get the latest available value that has
been measured.
Command Syntax
FETCh[:SCALar]:HARMonics:POWer[:ACTive]:FUNDamental?
Returned value
<NR2>
MEASure[:SCALar]:HARMonics:POWer[:ACTive]:FUN
Damental?
This command will trigger the measurement operation to get a real-time power
fundamental harmonic.
Command Syntax
MEAS[:SCALar]:HARMonics:POWer[:ACTive]:FUNDamental?
Returned value
<NR2>
FETCh[:SCALar]:HARMonics:POWer[:ACTive]:THARm
onic?
This command is used to read the latest pre-processing total harmonic power.It
will not affect the device setting values and will not trigger a new measurement
operation.It only requires to get the latest available value that has been
measured.
Command Syntax
FETCh[:SCALar]:HARMonics:POWer[:ACTive]:THARmonic?
Returned value
<NR2>
MEASure[:SCALar]:HARMonics:POWer[:ACTive]:THA
Rmonic?
This command will trigger the measurement operation to get a real-time total
harmonic power.
Command Syntax