Calibration Commands
Copyright © Itech Electronic Co., Ltd.
74
Relevant command
CAL:STAT CAL:SAV
CALibrate:VOLTage[:LEVel]
This command is only used in calibration mode. It enters a calibration voltage
value that you obtain by reading an external meter. You must first select a
calibration level (with CALibrate:VOLTage:POINt) for the value being
entered.These constants are not stored in nonvolatile memory until they are
saved with CALibrate:SAVE.
Command Syntax
CALibrate:VOLTage[:LEVel] <NRf>
Parameters
external reading
Unit
V (volts)
Example:CAL:VOLT 3.2223
Relevant Command
CAL:STAT CAL:SAV