Calibration Commands
Copyright © Itech Electronic Co., Ltd.
71
Chapter15 Calibration Commands
CALibration:ZERO
Zero calibration command.
Command Syntax
CALibration:ZERO
Parameters
None
CALibrate:SECure[:STATe]
This command is used to enable or disable calibration mode.
The calibration mode must be enabled before the load will accept any other
calibration commands.The first parameter specifies the enabled or disabled
state.The second parameter is the password.It is required if the calibration
mode is being enabled and the existing password is not 0. If the password is not
entered or is incorrect, an error is generated and the calibration mode remains
disabled.The
query
statement
returns
only
the
state,
not
the
password.Whenever the calibration state is changed from enabled to disabled,
any new calibration constants are lost unless they have been stored with
CALibrate:SAVE.
Command Syntax
CALibrate:SECure[:STATe] <bool> [,<SRD>]
Parameters
0 | 1 | OFF | ON [,<password>]
Reset value
ON
Example
CAL:SEC 0, N3301A CAL:SEC ON
Query Syntax
CALibrate:SECure[:STATe]?
Parameters
<NR1>
Relevant command
CAL:SAVE CAL:INIT
CALibrate:INITial
This command can only be used in calibration mode.It restore factory