CALCulate Commands
Copyright © Itech Electronic Co., Ltd.
64
Parameters
TIME|MAN|TINTerval
Query Syntax
INTegral:STOP:SOURce?
[:CALCulate]:INTegral:STOP[:IMMediate]
When integral stop mode is set to CMD,sending this command will stop the
integral operation.When stopped,integral value will not be cleared to zero.
Command Syntax
INTegral:STOP
Parameters
None
Query Syntax
None
[:CALCulate]:INTegral:STOP:DATE:<yy,MM,dd>
[:CALCulate]:INTegral:STOP:TIME: <hh,mm,ss>
This command is used to set the stop date,when time reached,device stops
integral operation.
Command Syntax
INTegral:STOP:DATE:<yy,MM,dd>
INTegral:STOP:TIME: <hh,mm,ss>
Parameters
<yy,MM,dd> <hh,mm,ss>
Query Syntax
INTegral:STOP:DATE?
INTegral:STOP:TIME?
[:CALCulate]:INTegral:STOP:TINTerval
<
hhhh,mm,ss >
When integral stop mode is set to TINTerval,user can set the integral measuring
period via this command.When time length reached,integral operation will be
terminated.
Command Syntax
INTegral:STOP:TINTerval < hhhh,mm,ss >
Parameters
hhhh,mm,ss