Sense Relevant Commands
Copyright © Itech Electronic Co., Ltd.
48
Parameters
OFF|ON
Query Syntax
CURRent:EXS2:STATe?
[:SENSe]CURRent:SRATio:EXS1 <NRf>
The command is used to set the conversion ratio of the external current sensor
1.
Command Syntax
CURRent:SRATio:EXS1 <NRf>
Parameters
0.001
~
9999.999
Query Syntax
CURRent:SRATio:EXS1?
[:SENSe]CURRent:SRATio:EXS2 <NRf>
The command is used to set the conversion ratio of the external current sensor
2.
Command Syntax
CURRent:SRATio:EXS2 <NRf>
Parameters
0.001
~
9999.999
Query Syntax
CURRent:SRATio:EXS2?
[SENSe:]CURRent:EXS:RANGe
This command is used to set the current range of external sensor.Default unit is
V.
Command Syntax
CURRent:EXS:RANGe <NRf>
Parameters
Model
Current range of external sensor
IT9121
CF=3: 2.5V, 5V, 10V, 50mV, 100mV, 200mV, 500mV, 1V, 2V
CF=6: 1.25V, 2.5V, 5V, 25mV, 50mV, 100mV, 250mV, 0.5V, 1V
IT9121E
CF=3: 2.5V, 5V, 10V, 50mV, 100mV, 200mV, 500mV, 1V, 2V
CF=6: 1.25V, 2.5V, 5V, 25mV, 50mV, 100mV, 250mV, 0.5V, 1V
IT9121H
CF=3: 2.5V, 5V, 10V, 50mV, 100mV, 200mV, 500mV, 1V, 2V