Technical
Specification
Copyright ©ITECH Electronic Co., Ltd.
40
Protection range
OPP
≒
310W
OCP
≒
3.3A
≒
16.5A
OVP
≒
530V
OTP
≒
85
℃
specification
Short circuit
current
(
CC
)
≒
3.3A
≒
16.5A
voltage
(
CV
)
0V
0V
resistance
(
CR
)
≒
240mΩ
Input
impedence
≒
500KΩ
External analog monitoring
I-Monitor
0~10V
Correspond
ing current
0~15A
Ac power supply
voltage
110V
220V
50/60Hz
Apparent
power
Max: 50VA
Fuse
specificat
ion
T2.5A
T1.25A
dimension
W*H*D (mm)
214.5mm*88.2mm*354.6mm
Weight
( net)
5Kg
Storage
temperatur
e
-20
℃
~70
℃
*1
The voltage/current input is no less than 10%FS (FS indicates the full scope).
*2
Range of current readback value: (1/(1/R+(1/R)*0.2%+0.01),1/(1/R-(1/R)*0.2%-0.01)
a) When voltage input value is less than 10%FS: 0.2%+0.1/Vin (s);
b) When current input value is less than 10%FS, the load current precision is:
±(0.2%xVin/R3mA);
*3
Range of resistance readback value: ( 1/(1/R+(1/R)*0.2%+0.001),1/(1/R-(1/R)*0.2%-0.001)
a) When voltage input value is less than 10%FS: 0.2%+0.05/Vin (s);
b) When current input value is less than 10%FS, the load current precision is: ±
(
0.2%xVin/R10mA
)
;
*4
The voltage/current input is no less than10%FS
*5
Increasing/decreasing slope: 10%~90% increasing slope of current when the current is from 0
to maximum current.
*6
Minimum rise time: 10%-90% current rise time
*The above specifications may be subject to change without prior notice.
4.2 Supplementary parameters
Memory capacity:100 groups
Suggested calibration frequency
:
1time/year
Input levels of AC power supply (which can be selected through the switch at
load bottom)
Option Opt.1: 220V ±10% 50Hz/60Hz