Function
and Features
Copyright ©ITECH Electronic Co., Ltd.
13
Fig. 3-1 Voltage-Current Relation Schema under CC Mode
Under CC mode, the electronic load provides three ways to set constant
current.
⚫
Rotate adjusting knob to set constant current value.
⚫
Use numeric keys to input current value and press
[Enter]
key to confirm
set constant current value.
⚫
Use
moving cursor and press
to adjust values at
corresponding positions.
Operation steps
1. Press
[CC]
key and
[Shift]
+
[CV]
to enter parameter setting screen.
Constant Current
Range=0.000A
2. Set maximum working current value and press
[Enter]
for confirmation.
Constant Current
Range =1.000A
▼
3. Set maximum voltage value and press
[Enter]
key.
Constant Current
High=500V
▼
4. Set minimum voltage value and press
[Enter]
key.
Constant Current
Low=0.1V
▼
5. Set high and low rate and press
[Enter]
key.
Constant Current
High-Rate Low-Rate
▼
6. Set ascending slope and press
[Enter]
key.
Constant Current
Rise up=0.001A/uS
▼
7. Set descending slope and press
[Enter]
key.
Constant Current
Rise down=0.001A/uS
8. Complete parameter setting.
10.0000V 0.0000A
0.00W CC=1.000A
Load voltage
I
V
Set current
CC mode
Load current