Function
and Features
Copyright ©ITECH Electronic Co., Ltd.
23
Taking CC mode as example (operations under other modes are similar), when
output voltage and current of the tested instrument are 10 V and 3 A
respectively, the load current will switch from 1 A and 2 A. Set the dynamic test
parameters and perform the test as follows:
1. Press
[Shift]
+ 2 keys.
TRANSITION
On Off
2. Operate
key and move to On. Press
[Enter]
key and select
Continuous. Press
[Enter]
key (The Trig lamp that indicates VFD screen
status is on)
TRANSITION
Continuous Pulse Toggle
3. Operate
key and select high range and low range. Move to the
High-Rate and press
[Enter]
key.
TRANSITION
High-Rate Low-Rate
4. Set ascending slope and press
[Enter]
key.
TRANSITION
Rise up=5.000A/uS
5. Set descending slope and press
[Enter]
key.
TRANSITION
Rise down=5.000A/uS
6. Set A value and press
[Enter]
key.
TRANSITION
Level A=1A
7. Set B value and press
[Enter]
key.
TRANSITION
Level B=2A
8. Set frequency value and press
[Enter]
key.
TRANSITION
Frequnce=50Hz
(
0.01-25000Hz
)
9. Set duty ratio and press
[Enter]
key.
TRANSITION
Duty=98%
(
0.1%-99.9%
)
10. Start dynamic test and operate
key. Move to On and press
[Enter]
key.
TRANSITION
On Off
T
Continuous Transient Operation